The RadonTec Radona , is the new way measure your radon air level. The device not only 
impresses  with  its  attractive  appearance,  but  also  with  its  inner  values.  Thanks  to  the  FTLAB 
ionization  chamber  technology  used  by  Waltec,  the  Radona  radon  measuring  instrument  can 

achieve up to 20 times faster measuring times than 
other  comparable  measuring  instruments.  In 
addition,  the  Radona  can  also  be  used  for  time-
resolved measurements, as the radon concentration 
is measured every 10 minutes. The radon measuring 
instrument  Radona  was  developed  especially  for 
professional  users  and  interested  private  users.  In 
addition  to  the  radon  meter  itself,  the  scope  of 
delivery  includes  a  power  supply  unit  and  a  USB-C 
cable  (for  reading  the  data  with  the  PC  and  for 
charging).  With  the  graphical  display  you  have 
different  views  on  the  measured  data  even  without 
transferring the data to a personal computer.  

 

 

Содержание Expert+

Страница 1: ...g times than other comparable measuring instruments In addition the Radona can also be used for time resolved measurements as the radon concentration is measured every 10 minutes The radon measuring instrument Radona was developed especially for professional users and interested private users In addition to the radon meter itself the scope of delivery includes a power supply unit and a USB C cable...

Страница 2: ... weight 840 g connectivity USB USB C Data storage 10 minutes or 60 minutes interval adjustable of radon date time air pressure temperature humidity ring memory older data are overwritten by the new ones Data storage 2 months at 10 minute intervals 12 months in 60 minutes interval Display LCD full graphic with different views Battery lasts for approx 24 hours device can be charged continuously oper...

Страница 3: ...Allows scrolling down function exists only in some views 2 Use this key to navigate through the individual views 3 To switch on press the button briefly To switch off press and hold for approx 3 seconds until Off appears on the screen ...

Страница 4: ...tween 5 10 minutes 2 Current Live radon value 3 WLAN symbol crossed out as here in the picture means that the device does not have a WLAN module 4 Indicates the battery level When the device is being charged a charging symbol appears 5 current humidity sensor is queried every 10 minutes 6 current air pressure sensor is queried every 10 minutes You can access the different views by pressing this bu...

Страница 5: ...4 The hook means that the measuring chamber is functioning correctly 049 would indicate a problem with the measuring chamber 5 battery level 6 Indicates the tendency of radon development Arrow up means that the radon value tends to rise arrow horizontal as in the figure indicates a stable radon value Arrow down predicts a decreasing radon value 7 current air pressure 8 current humidity ...

Страница 6: ...ifferent views 1 Indicates the running time of the measuring device since the last reset 2 Displays the Radon averages of the last day the last 7 days the last month and the last year 3 The statistically calculated inaccuracy is always displayed behind the mean values ...

Страница 7: ... the hours Y axis is the height of the radon value The Y axis adapts to the highest radon value On the far right there is always the bar which shows the current value marked by an unfilled bar An unfilled bar can still change in height until it becomes a filled bar ...

Страница 8: ... always rightmost The graph builds up from right to left and is continuous after 90 hours With this view you always have the radon concentration of the last 2 days at a glance Graphical bar display of the radon value of the last 31 days On the right is the current day ...

Страница 9: ...9 Graphical bar display of the radon value at month level On the right is the current month this bar changes until the month measurement is completed and another bar is displayed ...

Страница 10: ... meter This time is also used when saving the measured values 4 With this menu item all data in the meter can be deleted reset To do this press the middle button for approx 5 seconds 5 This menu item can be used to switch off the meter alternatively by pressing the On Off button on the meter for about 4 seconds 6 Interval 1 hour or 10 minutes The setting can be adjusted in the Radona Data Download...

Страница 11: ...11 Menu items 4 and 5 can be selected with this key This action can then be performed by pressing the center button However to reset all data the center button must be held down for approx 5 seconds ...

Страница 12: ...of charge The current version is enclosed with the measuring device on a USB stick The above package already includes the USB driver which can also be downloaded directly from the manufacturer if required https www silabs com products development tools software usb to uart bridge vcp drivers ...

Страница 13: ...mbered this appears in the export file and when downloading the data in the measuring instrument 7 Contrast of the screen can be adjusted 8 With this key the screen and the keys of the measuring instrument are locked for measurements at the customer Charging and measuring continue as usual 9 Keys on the meter are disabled 10 This bar can be moved with the left mouse button to select the zoom area ...