Introduction
This topic describes additional information pertinent to the use of the ABL77
analyzer.
Suggested tools
•
The following tools are recommended for various service procedures:
•
A
Sample Path Obstruction Tools
(
REF: 905-674
) kit can assist in removing
fluid path obstructions.
•
A
Zeroing Cassette
(
REF: 902-578
) is used to zero the sensor channels
•
A
Valve Board Calibration Cassette
(
REF: 920-729
) is used to calibrate the
Hct and SC channels between the valve board and analog board.
Warranty
sensor cassette
returns
Any sensor cassette that fails during use should be returned via the distributor to
SenDx Medical, Inc. Please return all the following items for full warranty credit
and to assist in a complete failure investigation:
•
Sensor cassette
•
Foil Pouch
•
Printout of the failed Event Record. Only the Event Record contains a complete
set of information important for failure investigation. An Event Record is printed
from the
Recall / Event
screen.
•
Highlight the failed calibration, initialization or QC analysis
•
Press the
icon to print an Event Record of this failed event.
Analyzer
component
returns
Any analyzer component returned for credit or for failure investigation should be
accompanied by a copy of the service work order. Please be as complete as
possible in describing the events that led up to the failure, any troubleshooting
steps taken, and resolution of the problem. Please include the serial number of the
analyzer repaired.
Cassette pump
tubing
During calibrant pump cycles,
while the roller wheel is
spinning, the cassette pump
tubing can be "snapped"
repeatedly to help dislodge any
air bubbles residing in the
cassette tubing (see Figure 3-1).
This simple step can alleviate a
number of the more common
calibration and initialization
failure experiences.
Figure
3-1
3-4
Содержание ABL 77 Series
Страница 19: ...Chapter 2 Analyzer Description ABL77 Service Manual This page intentionally left blank 2 4 Rev F ...
Страница 26: ...ABL77 Service Manual Chapter 2 Analyzer Description This page intentionally left blank Rev F 2 11 ...
Страница 40: ...ABL77 Service Manual Chapter 2 Analyzer Description This page intentionally left blank Rev F 2 25 ...
Страница 227: ...Chapter 6 Replacements ABL77 Service Manual This page intentionally left blank 6 32 Rev F ...
Страница 333: ...ABL77 Service Manual Chapter 9 Spare Parts Modules and Externals Continued Module diagrams continued Rev F 9 3 ...
Страница 335: ...ABL77 Service Manual Chapter 9 Spare Parts Upper Module Components Continued Upper module component diagram Rev F 9 5 ...
Страница 344: ...Chapter 9 Spare Parts ABL77 Service Manual This page intentionally left blank 9 14 Rev F ...