Chapter 6: Replacements
ABL77 Service Manual
Touch Screen Cable
Touch screen
cable
Follow these steps to replace the touch screen cable.
Step Action
1.
Remove the electronics module. One end of the touch screen cable
will be disconnected when the upper module is removed.
2.
Disconnect the interface board from the CPU board. It is not necessary
to disconnect the interface board cables.
3.
Disconnect the old touch screen cable from the front of the touch
screen controller board.
4.
Clean the old RTV adhesive from the header on the touch screen
controller board.
5.
Connect the new touch screen cable (item
1
in figure 6-25) to the
16-pin header on the front of the touch screen controller board.
Position the connector in the bottom row on the far right side so that
pin 1 (different color strip) is to the right.
Figure 6-25
6.
Place RTV adhesive on the ends/edges of the connections between the
touch screen controller board and the touch screen cable.
NOTE:
Do not get RTV adhesive on the pins.
7.
Re-assemble the analyzer starting with the electronics module.
6-42
Rev.
F
Содержание ABL 77 Series
Страница 19: ...Chapter 2 Analyzer Description ABL77 Service Manual This page intentionally left blank 2 4 Rev F ...
Страница 26: ...ABL77 Service Manual Chapter 2 Analyzer Description This page intentionally left blank Rev F 2 11 ...
Страница 40: ...ABL77 Service Manual Chapter 2 Analyzer Description This page intentionally left blank Rev F 2 25 ...
Страница 227: ...Chapter 6 Replacements ABL77 Service Manual This page intentionally left blank 6 32 Rev F ...
Страница 333: ...ABL77 Service Manual Chapter 9 Spare Parts Modules and Externals Continued Module diagrams continued Rev F 9 3 ...
Страница 335: ...ABL77 Service Manual Chapter 9 Spare Parts Upper Module Components Continued Upper module component diagram Rev F 9 5 ...
Страница 344: ...Chapter 9 Spare Parts ABL77 Service Manual This page intentionally left blank 9 14 Rev F ...