TitraLab 870 Reference Manual
Page 98
Display
measurement
Press
5
in the Electrode window.
The signal measured of a connected electrode in the current system
is displayed. If several electrodes are connected, select the electrode
at the
ID
line.
Depending on the type of electrode connected, the display shows:
•
pH and corresponding potential difference in mV (pH electrodes)
•
potential difference in mV (metal/redox or non-calibrated ISE
electrodes)
•
concentration in the electrode calibration unit (calibrated ISE
electrodes)
•
temperature in °C (temperature sensors)
•
potential difference in mV and polarisation current in µA (mV(i>0)
measurement mode for a double metal electrode). The polarisation
signal parameters are entered in the Method parameters or
Reagent parameters menus.
Refer to "Applied signal (AC/DC)", page 52
.
Refer to "Current value", page 88
.
•
conductivity measured in mS/cm at sample temperature (conduc-
tivity cells). If the conductivity is not calibrated, the instrument dis-
plays a conductivity with a cell constant value equal to 1 cm
-1
.
The sample temperature is measured or is equal to 25 °C. There is
no temperature correction performed.
To get accurate measurements, it is therefore recommended to
calibrate the conductivity cell at a given temperature and
thermostat the sample to that temperature before running the
measurement.
Press
1
to apply or stop stirring.
Press
Esc
to stop measurements.
Содержание TitraLab 870
Страница 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Страница 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Страница 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Страница 23: ...Practical examples s...
Страница 24: ......
Страница 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Страница 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Страница 41: ...Glossary...
Страница 42: ......
Страница 127: ...TitraLab TitraLab 870 Reference Manual Page 127 ERR 32 SAC error Refer to SAC switch Off On SAC error page 219...
Страница 255: ...Appendixes...
Страница 256: ......
Страница 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Страница 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Страница 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Страница 286: ...TitraLab TitraLab 870 Reference Manual Page 286...