Titra
Lab
®
TitraLab 870 Reference Manual
Page 243
Stop point
In Continuous IP, Monotonic IP or Dynamic IP method, the Stop point
is the potential (or pH) at which the titration will stop if the Maximum
volume or the selected Number of IP has not been reached.
Select the Stop point on the other side of the expected inflection
point(s).
Set the Stop point with respect to the Direction parameter.
Enter in:
Method parameters menu (Monotonic IP, Dynamic IP or Continuous
IP).
Reagent parameters menu (Monotonic IP, Dynamic IP or Continuous
IP).
Range available:
-9.000 pH to +23.000 pH (or -2000 mV to +2000 mV)