Titra
Lab
®
TitraLab 870 Reference Manual
Page 201
Reprocessing a
curve (rules)
The so-called “Reprocess“ method must satisfy a few conditions
regarding the original method the curve comes from.
1.
Same electrode and reagent systems,
2.
Same sample preparation (diluted or non-diluted samples),
3.
Same type of result unit (volume or mass units),
4.
Same type of titration (Direct titration or Back titration),
5.
Conditions on the titration mode :
EP or IP mode for a curve obtained in IP,
EP only for a curve obtained in EP.
6.
For a reagent calibration method: same reagent ID.
In Automatic mode, it is recommended to create the so-called
“Reprocess“ method by copying the titration method the curve
comes from. Then you have just to change the desired
parameters.
When editing the so-called “Reprocess“ method, you can change the
following parameters (the other parameters are useless if changed or
can lead to errors while running the curve reprocessing):
End point method:
Edit method menu: Mode (with condition no.5, see above),
Parameters menu: Direction, Number of EP, Value of EP,
Sample menu: no change possible,
Results and Printouts menus: All parameters can be changed (with
condition no.3, see above),
QC Data menu: Minimum value and Maximum value for each result.
Continuous IP, monotonic IP and dynamic IP methods:
Edit method menu: Mode (with condition no.5, see above),
Parameters menu: Direction, IP Filter, IP Reject, Number of IP,
Minimum and Maximum ordinates for each IP,
Sample menu: no change possible,
Results and Printouts menus: All parameters can be changed (with
condition no.3, see above),
QC Data menu: Minimum value and Maximum value for each result.
Содержание TitraLab 870
Страница 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Страница 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Страница 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Страница 23: ...Practical examples s...
Страница 24: ......
Страница 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Страница 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Страница 41: ...Glossary...
Страница 42: ......
Страница 127: ...TitraLab TitraLab 870 Reference Manual Page 127 ERR 32 SAC error Refer to SAC switch Off On SAC error page 219...
Страница 255: ...Appendixes...
Страница 256: ......
Страница 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Страница 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Страница 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Страница 286: ...TitraLab TitraLab 870 Reference Manual Page 286...