TitraLab 870 Reference Manual
Page 183
QC data menu
This menu is available for measurement and titration methods.
To access:
1.
Enter the Main window.
2.
Select or create a method.
3.
Select
QC sample = Yes
in the Edit method menu.
4.
Use the
LEFT/RIGHT
arrow keys to move to the last Edit method
display.
5.
Press
5
QC data.
QC ID
Enter the name of the QC sample (16 alphanumeric characters). This
name is called up when you run a QC sample analysis.
QC not required
Message appears at the start of a sequence, that originally included a
method programmed with a QC sample.
The method to has now been reprogrammed without QC sample.
Go to Sequence/Sample stack, Edit sequence screen and remove the
QC sample from the sequence.
QC periodicity
elapsed
This message is displayed at the start of a method requiring a QC
sample. The Periodicity, entered in the QC Data screen of the method
has elapsed. For example, if Periodicity = 10 samples, then a QC
sample must be performed every 10 samples.
Press
✓
and run a QC analysis.
Содержание TitraLab 870
Страница 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Страница 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Страница 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Страница 23: ...Practical examples s...
Страница 24: ......
Страница 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Страница 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Страница 41: ...Glossary...
Страница 42: ......
Страница 127: ...TitraLab TitraLab 870 Reference Manual Page 127 ERR 32 SAC error Refer to SAC switch Off On SAC error page 219...
Страница 255: ...Appendixes...
Страница 256: ......
Страница 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Страница 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Страница 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Страница 286: ...TitraLab TitraLab 870 Reference Manual Page 286...