Titra
Lab
®
TitraLab 870 Reference Manual
Page 133
Global flush
burettes
A “Global flush” function is available to expell the air bubbles that may
be trapped in the tubings. All installed burettes are flushed
simultaneously.
Perform the following operations:
1.
Display the Burette functions menu:
2.
Select
all
for
Address
.
3.
Press
3
Global flush.
The “Global flush” procedure runs as follows:
•
Empty 4 ml or the full burette (if burette volume < 4 ml),
•
Fill,
•
Empty 10 ml or the full burette (if burette volume < 10 ml),
•
Fill.
Global variable
A global variable is used to transfer a result from one method to
another. To assign a global variable to a result, select Gi (i = 1 to 10).
This global variable can be inserted into the equation used to
calculate the method result(s). Each of the 10 global variables can be
visualised in the GLP - Archives menu.
Never use these variables to transfer a result inside a same
method. However, you can use a global variable to transfer a
result from one method to another method of a same Coupled
method.
Enter in:
Edit method > Results
Содержание TitraLab 870
Страница 1: ...TitraLab 870 pH EP IP EC ISE Titration Workstation Reference Manual D21T060...
Страница 2: ...D21T060 Printed by Radiometer Analytical SAS 2008 02F...
Страница 22: ...TitraLab TitraLab 870 Reference Manual Page 22...
Страница 23: ...Practical examples s...
Страница 24: ......
Страница 32: ...TitraLab TitraLab 870 Reference Manual Page 32...
Страница 40: ...TitraLab TitraLab 870 Reference Manual Page 40...
Страница 41: ...Glossary...
Страница 42: ......
Страница 127: ...TitraLab TitraLab 870 Reference Manual Page 127 ERR 32 SAC error Refer to SAC switch Off On SAC error page 219...
Страница 255: ...Appendixes...
Страница 256: ......
Страница 258: ...TitraLab TitraLab 870 Reference Manual Page 258...
Страница 262: ...TitraLab TitraLab 870 Reference Manual Page 262...
Страница 280: ...TitraLab TitraLab 870 Reference Manual Page 280...
Страница 286: ...TitraLab TitraLab 870 Reference Manual Page 286...