8.3 Service
8.3.1 Service
The Bantam Plus equipment is designed for testing meters and used as part of your
organization's quality control system. The procedure below outlines the limited parts
that are customer replaceable.
Over extended use, the current carrying socket jaws may foul with contaminates
such as oil or grease and no longer produce the contact necessary to operate to
specification. The following procedure describes how to periodically clean or
replace these jaws.
Step 1. Ensure the Bantam Plus is switched off and isolated from line voltage.
Step 2. Remove the five Torx drive screws from the socket cover.
Step 3. Remove the socket cover to expose the socket assembly.
Step 4. Note that each current conductor has a replaceable liner. Clean each liner
using compressed air being careful that all loose debris is removed. Should oil be
present remove the liner from the Bantam Plus before using an oil solvent for
cleaning.
Remove the five Torx
screws to expose the
socket assembly