32
RIDA qLine® autoBlot | User manual
Fig. 23:
Autoblot protocolversion
The AutoBlotProtocolversion may not be changed. The AutoBlotSoftwareversion
corresponds with the Application version and the AutoBlotFirmwareversion corresponds
with the FirmwareVersion shown in the Information tab (see section 6.1.1).
Make sure that the device is connected to the power supply during all update processes.
6.3.2 Full automatic processing
When the RIDA qLine® Allergy tests are processed fully automatically in RIDA qLine®
autoBlot, the bar codes of the test tubes will be scanned and sent to the connected RIDA
qLine® Soft program. When RIDA qLine® Soft is connected to the LIS, all requests will be
retrieved from the LIS, and a worklist will be created automatically. This worklist will be
sent automatically to RIDA qLine® autoBlot and displayed on the touchscreen.
If RIDA qLine® Soft is not connected to a LIS, the test requests of the specimens to be
tested must be entered manually in RIDA qLine® Soft in advance. Once the specimens are
scanned in the machine, a worklist is created, sent to the machine, and appears on the
RIDA qLine® autoBlot screen.
The lab IDs that were entered in the program and the lab IDs on the test
tubes must be identical. If the manually entered lab IDs are not identical
to the scanned lab IDs, a worklist cannot be created.