Q150T Sample Preparation System
Q150T - Instruction Manual
79
10473 - Issue 5
Table 25
Carbon Pulse/Rod Evaporation profile parameters
Name
Default
Value
Minimum
Value
Maximum
Value
Comment
Material
Carbon
Fixed; read only parameter
Pulse current (A)
60 (cord)
75 (rod)
0
120
Current applied to cord during
the evaporation pulse
Max Pulse
Voltage (V)*
20 (cord)
10 (rod)
0
20
The voltage across the cord is
limited to this value during the
evaporation pulse
Pulses Length (s)
1
0.02
10
Time is second that the pulse is
applied for
Number of pulses
1
1
5
Number of consecutive pulse
cycles
Pulse interval (s)
10
1
30
The time between the start of
consecutive pulse cycles. Must
exceed pulse length.
Outgas time (s)
10
2
60
Time cord or rod is outgassed
before pulse cycle(s).
Outgas current
(A)
30 (cord)
20 (rod)
0
40
Current applied at the start of
outgas step.
Max Outgas
current (A)
40
0
50
Maximum current permitted
during outgas.
Max Outgas
voltage (V)
10
0
20
Maximum voltage permitted
during outgas
Outgas source*
Yes
No
Yes
Include or exclude outgassing
step
Tooling factor
5
0.1
10
Tooling factor for this process
Pump hold*
No
No
Yes
If
Yes,
Q150T pumps down to
outgas-ready state and waits for
user input.
Note:
If hold time (set in
System | Vacuum | Pump hold
time
– see page 46 is
exceeded, Q150T vents
chamber.
Flush chamber*
No
No
Yes
Set to
Yes
to flush chamber
with nitrogen before pumping
down.
* Admin group level required to edit this property