Q150T Sample Preparation System
Q150T - Instruction Manual
15
10473 - Issue 5
2.3.4
Vacuum Connections
1.
If you are using an existing or alternative vacuum pump, and have any difficulty
with connections, please ask for advice.
2.
If the instrument is NOT going to be vented into an extraction system, fit an Oil
Mist Filter with metal adapter to the outlet of the vacuum pump. Please refer to
the manufacturer’s instructions.
3.
Ensure that the vacuum pump is filled with the correct oil.
4.
If the vacuum pump is fitted with an on/off switch, ensure that it is set to the 'on'
position to allow the instrument to control its power supply.
5.
Connect the flexible vacuum hose to the spigot on the rear panel using the hose
clip provided. This is a push-on fit to the Instrument. Ensure that this is firmly in
place to the full length of the spigot.
A rotating vacuum spigot is available for more convenient hose
connection in installations where the instrument is to be located close
to a back wall (see page 69).