0.001-6 GHz Single Channel Upconverter
IEEE 488.2 Command Reference
6.10 *SDS
<
NRf
>
The
*SDS
(Save Default device Settings) command restores the specified state of the QM1013-101
0.001-6 GHz Single Channel Upconverter to the default state settings. There are 5 re-writeable memory
locations, specified by choosing an index between 1 and 5. The parameters restored are the
RF Power
State, External LO1 Enabled, External LO1 Switch Override, External LO2 Enabled, External LO2 Switch
Override, External Reference Enabled, External Referece Switch Override, Tune Frequency, Attenuator
1 Setting, Attenuator 2 Setting, and Attenuator 3 Setting
. For a detailed description of the parameters
saved, see the relevent command descriptions in this manual. Issuing this command does not change
the current settings. NOTE: The MEM_CLR button on the back panel will rewrite the contents of states
1-5 with the contents of state 0 when depressed, serving as a reset to factory defaults.
Syntax
*
SDS
Space
NRf
Allowed Values
The
NRf
parameter can be any integer in the range of 1 to 5.
Error Message
If the parameter is not in the recognized format, error
-102, "Syntax error"
occurs.
If the parameter is less than 1 or greater than 5, error
-222, "Data out of range"
occurs.
Quonset Microwave
Revision 1.0.1
74