Combined RF Upconverter/Downconverter
IEEE 488.2 Command Reference
6.9 *SAV
<
NRf
>
The
*SAV
(SAVe) command saves the current setup to non-volatile memory. There are 5 re-writeable
memory locations, specified by choosing an index between 1 and 5. Any of the 5 states can be restored
on startup when specified with the SYSTem:BOOTstate command. The parameters saved are the
RF
Power State, Upconverter LNA Power State, Reference Frequency, External Reference Enable, External
Reference Switch Override, Tune Frequency, External LO1 Enable, External LO1 Switch Override, LO1
PLL Mode, LO1 Reference Divider, External LO2 Enable, External LO2 Switch Override, Upconverter
RF Attenuation, Downconverter RF Attenuation
. For a detailed description of the parameters saved, see
the relevent command descriptions in this manual. NOTE: The MEM_CLR button on the back panel will
rewrite the contents of states 1-5 with the contents of state 0 when depressed, serving as a reset to
factory defaults.
Syntax
*
SAV
Space
NRf
Allowed Values
The
NRf
parameter can be any integer in the range of 1 to 5.
Error Message
If the parameter is not in the recognized format, error
-102, "Syntax error"
occurs.
If the parameter is less than 1 or greater than 5, error
-222, "Data out of range"
occurs.
Quonset Microwave
Revision 2.0.0
70