3-8 Troubleshooting
MedTrack CR60 Rehabilitation Treadmill Service Manual
*The type and the rate of bearing noise can help determine which bearing
is at fault. The number of knocks per minute varies with treadmill speed
because the roller speeds change as the belt speed changes.
**Use a mechanical stethoscope or a piece of hose about two feet long, to
isolate bearing problems. Hold one end of the hose near the suspected
bearing and the other end near your ear. Compare several bearings to
determine which is the faulty one.
Flow Charts
Use the flow charts on the following pages for step-by-step
troubleshooting.
Popping during grade
change.
Faulty elevation chain
alignment.
Adjust alignment of
sprockets.
Noise
Possible Cause
Action
Error Code Flow Chart
Page
EPHI and EPLO
“Bus Voltage Flow Chart” on page 3-9.
E201
“Grade Error Flow Chart” on page 3-10.
E203 (Drive Motor)
“Drive Motor Error Flow Chart” on page 3-11.
E204
“Communications Link” on page 3-12.
Grade Problems
Содержание MedTrack CR60
Страница 1: ...MedTrack CR60 Rehabilitation Treadmill Service Manual P N 042101 001 Rev A...
Страница 4: ......
Страница 27: ...MedTrack CR60 Rehabilitation Treadmill Service Manual Troubleshooting 3 13 Grade Problems Flow Chart...
Страница 28: ......
Страница 58: ......
Страница 62: ......
Страница 70: ......
Страница 78: ...C 4 Part Numbers MedTrack CR60 Rehabilitation Treadmill Service Manual 13 18 27 27 26 26 26 15 15...
Страница 80: ......
Страница 88: ......
Страница 89: ...E 3 Assembly Drawings Cardiac Stress Treadmills Service Manual Wiring Diagram Low Voltage Treadmill...
Страница 90: ...E 4 Assembly Drawings Cardiac Stress Treadmills Service Manual Wiring Diagram High Voltage Treadmill...