MPS 3 System – System Installation
5.5. Installing the MPS 3 System
5.5.1. Installing the MPS 3 Console
The MPS 3 System should be integrated with the Heart-Lung Machine (HLM) for
delivery of cardioplegia solution to the patient. This can be accomplished by
mounting the MPS 3 Console on the base of the HLM or on a Quest Medical supplied
MPS 3 Console Pole Mount. The MPS 3 Console should be placed in a location that
is convenient for safe and effective operation.
A Heart Lung Machine Adapter may be required to ensure safe and effective
installation of the MPS 3 Console prior to attempting to installation. Consult Quest
Medical Customer Service or Sales Representative to determine if an adapter is
required for installation.
For installation of the MPS 3 Console on the HLM Base, follow these steps:
1. Consult the HLM Operation Instructions prior to attempting to install the MPS 3
Console to ensure the console can be safely installed.
2. Once a placement location has been identified, clean and tidy the area. Ensure
the surface is
DRY, STABLE, and LEVEL
, prior to placing the MPS 3 Console.
3. If a Heart Lung Machine Adapter is required, consult the installation instructions
for the adapter and install the adapter.
Before installing the MPS 3 Console, consult the HLM
Operations Manual / User Manual to verify the MPS 3
Console can safely be installed on the HLM according to
recommended load restrictions. The HLM should be
configured to best fit the MPS 3 Console. Prior to
installation of the MPS 3 Console, clean and tidy the area /
surface in which the Console will be placed. Ensure the area
is clean, dry, stable and level prior to attempting to place the
MPS 3 Console. Never install an MPS 3 Console on a
surface that is not suitable to support the Console.
Avoid OPERATOR INJURY. Only qualified technicians
should attempt to install the MPS 3 Console. Use proper
lifting techniques. Ensure that all fingers are free from pinch
/ crush hazards while installing the MPS 3 Console.
Use the proper electrical connections. The MPS 3 console
power cord must be plugged into a dedicated hospital-
grade electrical outlet. The hospital biomedical engineering
department should periodically measure console ground
leakage current to verify acceptable limits. Use only Quest
Medical power cords.
In the event of AC Power loss or failure, the MPS 3 System
Battery Backup will be activated. When this occurs, the
console heaters will be disabled.
9
Содержание MPS 3
Страница 1: ...Quest Medical Inc...
Страница 11: ...MPS 3 System Introduction TTBD Time to be Delivered Term Definition Description 5...
Страница 12: ...MPS 3 System Introduction 6...
Страница 30: ...stem System Specifications and Performance This page intentionally left blank 8...
Страница 42: ...MPS 3 System Warnings and Cautions This page intentionally left blank 12...
Страница 60: ...MPS 3 System System Installation 18...
Страница 64: ...MPS 3 System System Installation Figure 21 Tighten Pole Clamp to Secure Mounting Arm 22...
Страница 90: ...MPS 3 System Startup This page intentionally left blank 9...
Страница 123: ...MPS 3 System Graphical User Interface GUI Overview Figure 80 Auto Mode Screen 33...
Страница 135: ...MPS 3 System Disposable Installation Figure 96 Align the Additive Cartridge 12...
Страница 149: ...MPS 3 System Functional Use Figure 108 New Case Parameters Page 2 Figure 109 New Case Parameters Page 3 3...
Страница 150: ...MPS 3 System Functional Use Figure 110 New Case Parameters Page 4 Figure 111 New Case Parameters Page 5 4...
Страница 151: ...MPS 3 System Functional Use Figure 112 New Case Parameters Page 6 Figure 113 New Case Parameters Page 7 5...
Страница 152: ...MPS 3 System Functional Use Figure 114 New Case Setup Parameters Page 8 6...
Страница 167: ...MPS 3 System Functional Use Figure 130 Dose History Screen 21...
Страница 171: ...MPS 3 System Functional Use Figure 133 Case Report 1 of 4 Figure 134 Case Report Procedure View 25...
Страница 175: ...MPS 3 System Functional Use Figure 136 Case Report 2 of 4 29...
Страница 178: ...MPS 3 System Functional Use 32...
Страница 185: ...MPS 3 System Functional Use Figure 149 Error Report 39...
Страница 187: ...MPS 3 System Functional Use Figure 150 Entering Simulator Mode Warning Figure 151 Simulator Mode 41...
Страница 231: ...MPS 3 System Troubleshooting 40...
Страница 232: ...MPS 3 System Troubleshooting Figure 157 Alarm 314 Figure 158 Alarm 315 41...
Страница 233: ...MPS 3 System Troubleshooting This page intentionally left blank 42...
Страница 234: ...MPS 3 System Troubleshooting 43...
Страница 247: ...MPS 3 System System Maintenance Figure 165 Data Logs Screen 13...
Страница 252: ...MPS 3 System Service 4...
Страница 256: ...MPS 3 System Part Numbers and Accessories 4...
Страница 258: ...MPS 3 System Part Numbers and Accessories 6...
Страница 259: ...MPS 3 System Part Numbers and Accessories 7...
Страница 260: ......
Страница 268: ...MPS 3 System Appendix This page intentionally left blank 8...