MPS 3 System – System Maintenance
12.1.3.5.Changing the Water in the System
The water in the circulation system must be changed every 7 days. After changing,
add 25 ml of 3% Hydrogen Peroxide to the Hypothermic Reservoir and circulate to
inhibit microbial growth. After each use of the MPS 3 System, drain the Hypothermic
Reservoir to its drain spout, add an additional 25 ml of 3% hydrogen peroxide, to
compensate for dilution of melted ice, and circulate to ensure distribution throughout
the water circulation system.
Note: To ensure the hydrogen peroxide concentration is maintained above 100ppm
conduct daily monitoring of the concentration follows:
•
Drain 100 ml of water from the circulation system drain valve and discard.
•
Drain a minimum volume of 5 ml from the drain valve into a sterile sample
container for measurement and then discard the drain/adapter assembly.
•
Test the concentration of hydrogen peroxide using a peroxide test strip (The
MPS 3 disinfection procedure was validated using Indigo Instruments High
Level Peroxide Test Strip 33815-P400). Refer to test strip Manufacturer’s IFU for
instruction on using test strip.
•
If the H
2
O
2
concentration falls below 100ppm, repeat the Disinfection process.
12.1.4.System Controller Cleaning
The Controller display panel has a special coating that can be ruined by strong
cleaners. Do not spray water or other liquids directly on the screen or the case.
Protect the Controller from spills and dripping OR solutions. Prevent any kind of
adhesive application to the surface of the display panel.
Do not use alcohol wipes or antimicrobial wipes to clean the display panel as it may
damage the screen.
If the Controller does not respond to the operators touch or is visibly soiled, the
display panel needs cleaning. Pause the screen to avoid inadvertent parameter
changes. Clean the display panel with only a damp soft cloth to prevent scratching.
Clean spills immediately after the occurrence of the spill.
Clean the case, Power / Data cable and mounting arm with antimicrobial wipes.
10
Содержание MPS 3
Страница 1: ...Quest Medical Inc...
Страница 11: ...MPS 3 System Introduction TTBD Time to be Delivered Term Definition Description 5...
Страница 12: ...MPS 3 System Introduction 6...
Страница 30: ...stem System Specifications and Performance This page intentionally left blank 8...
Страница 42: ...MPS 3 System Warnings and Cautions This page intentionally left blank 12...
Страница 60: ...MPS 3 System System Installation 18...
Страница 64: ...MPS 3 System System Installation Figure 21 Tighten Pole Clamp to Secure Mounting Arm 22...
Страница 90: ...MPS 3 System Startup This page intentionally left blank 9...
Страница 123: ...MPS 3 System Graphical User Interface GUI Overview Figure 80 Auto Mode Screen 33...
Страница 135: ...MPS 3 System Disposable Installation Figure 96 Align the Additive Cartridge 12...
Страница 149: ...MPS 3 System Functional Use Figure 108 New Case Parameters Page 2 Figure 109 New Case Parameters Page 3 3...
Страница 150: ...MPS 3 System Functional Use Figure 110 New Case Parameters Page 4 Figure 111 New Case Parameters Page 5 4...
Страница 151: ...MPS 3 System Functional Use Figure 112 New Case Parameters Page 6 Figure 113 New Case Parameters Page 7 5...
Страница 152: ...MPS 3 System Functional Use Figure 114 New Case Setup Parameters Page 8 6...
Страница 167: ...MPS 3 System Functional Use Figure 130 Dose History Screen 21...
Страница 171: ...MPS 3 System Functional Use Figure 133 Case Report 1 of 4 Figure 134 Case Report Procedure View 25...
Страница 175: ...MPS 3 System Functional Use Figure 136 Case Report 2 of 4 29...
Страница 178: ...MPS 3 System Functional Use 32...
Страница 185: ...MPS 3 System Functional Use Figure 149 Error Report 39...
Страница 187: ...MPS 3 System Functional Use Figure 150 Entering Simulator Mode Warning Figure 151 Simulator Mode 41...
Страница 231: ...MPS 3 System Troubleshooting 40...
Страница 232: ...MPS 3 System Troubleshooting Figure 157 Alarm 314 Figure 158 Alarm 315 41...
Страница 233: ...MPS 3 System Troubleshooting This page intentionally left blank 42...
Страница 234: ...MPS 3 System Troubleshooting 43...
Страница 247: ...MPS 3 System System Maintenance Figure 165 Data Logs Screen 13...
Страница 252: ...MPS 3 System Service 4...
Страница 256: ...MPS 3 System Part Numbers and Accessories 4...
Страница 258: ...MPS 3 System Part Numbers and Accessories 6...
Страница 259: ...MPS 3 System Part Numbers and Accessories 7...
Страница 260: ......
Страница 268: ...MPS 3 System Appendix This page intentionally left blank 8...