MPS 3 System – System Maintenance
12.1.3.Water Circulation System Cleaning
The MPS 3 System water circulation system must be
disinfected prior to its first use, while in operation, and
while in storage at a frequency of 28 days.
DO NOT use bleach or any chlorine based cleaning
solutions in the MPS 3 circulation system. Chlorine may
damage the heat exchanger, resulting in water to blood
leakage AND possible patient injury.
Other disinfectants with the same active ingredients may
contain different additives which can influence the material
compatibility. Thus the use of any other disinfectant is
explicitly not recommended.
Immediately remove from service any device that show
discoloration or cloudiness in the water hoses which may
indicate bacterial growth. Consult hospital infection control
for further action. Replace hoses that connect to external
devices annually at a minimum to minimize biofilm
formation.
Do not leave the MPS disposable Heat exchanger primed
with water for a prolonged time as this may damage the
heat exchanger, resulting in water to blood leakage AND
possible patient injury.
Use Minncare HD to clean and disinfect the MPS 3 console
water circulation system. Follow all manufacturer’s
recommendations for preparation, storage, and disposal of
disinfection solution. Rinse MPS 3 per this IFU after
disinfecting.
The cleaning mode may be accessed by selecting H2O CIRC
from the MENU screen.
Prior to cleaning the water circulation system, ensure that
the delivery set and accessories have been removed, the
hypothermic reservoir has been drained and the water
circuit adapter has been installed.
5
Содержание MPS 3
Страница 1: ...Quest Medical Inc...
Страница 11: ...MPS 3 System Introduction TTBD Time to be Delivered Term Definition Description 5...
Страница 12: ...MPS 3 System Introduction 6...
Страница 30: ...stem System Specifications and Performance This page intentionally left blank 8...
Страница 42: ...MPS 3 System Warnings and Cautions This page intentionally left blank 12...
Страница 60: ...MPS 3 System System Installation 18...
Страница 64: ...MPS 3 System System Installation Figure 21 Tighten Pole Clamp to Secure Mounting Arm 22...
Страница 90: ...MPS 3 System Startup This page intentionally left blank 9...
Страница 123: ...MPS 3 System Graphical User Interface GUI Overview Figure 80 Auto Mode Screen 33...
Страница 135: ...MPS 3 System Disposable Installation Figure 96 Align the Additive Cartridge 12...
Страница 149: ...MPS 3 System Functional Use Figure 108 New Case Parameters Page 2 Figure 109 New Case Parameters Page 3 3...
Страница 150: ...MPS 3 System Functional Use Figure 110 New Case Parameters Page 4 Figure 111 New Case Parameters Page 5 4...
Страница 151: ...MPS 3 System Functional Use Figure 112 New Case Parameters Page 6 Figure 113 New Case Parameters Page 7 5...
Страница 152: ...MPS 3 System Functional Use Figure 114 New Case Setup Parameters Page 8 6...
Страница 167: ...MPS 3 System Functional Use Figure 130 Dose History Screen 21...
Страница 171: ...MPS 3 System Functional Use Figure 133 Case Report 1 of 4 Figure 134 Case Report Procedure View 25...
Страница 175: ...MPS 3 System Functional Use Figure 136 Case Report 2 of 4 29...
Страница 178: ...MPS 3 System Functional Use 32...
Страница 185: ...MPS 3 System Functional Use Figure 149 Error Report 39...
Страница 187: ...MPS 3 System Functional Use Figure 150 Entering Simulator Mode Warning Figure 151 Simulator Mode 41...
Страница 231: ...MPS 3 System Troubleshooting 40...
Страница 232: ...MPS 3 System Troubleshooting Figure 157 Alarm 314 Figure 158 Alarm 315 41...
Страница 233: ...MPS 3 System Troubleshooting This page intentionally left blank 42...
Страница 234: ...MPS 3 System Troubleshooting 43...
Страница 247: ...MPS 3 System System Maintenance Figure 165 Data Logs Screen 13...
Страница 252: ...MPS 3 System Service 4...
Страница 256: ...MPS 3 System Part Numbers and Accessories 4...
Страница 258: ...MPS 3 System Part Numbers and Accessories 6...
Страница 259: ...MPS 3 System Part Numbers and Accessories 7...
Страница 260: ......
Страница 268: ...MPS 3 System Appendix This page intentionally left blank 8...