MPS 3 System – Troubleshooting
11.4.5.Delivery Temperature Reads More Than 5° Above Water
Temperature
•
It is likely that the IR temperature sensor behind the heat exchanger has a dirty
lens. Clean the lens with water and a soft cloth.
•
If wiping the sensor does not fix the issue, ensure that there is not a heat source
pointed at the HEX or making contact with it.
11.4.6.Delivery/Vent Valves Errors
•
The antegrade, retrograde, or vent valve may have debris or buildup in the
mechanism. Rinse the valve with warm water and actuate the valve manually
several times to clear the debris.
11.4.7.Low/No Water Flow in Circulation System
•
The circulation system may not be primed properly.
•
If using Hypothermic Reservoir:
o
Ensure the Hypothermic Reservoir has water filled to the drain
port.
o
Elevate the Hypothermic Reservoir above the MPS 3 System so
that the outlet port of the reservoir is above the heat exchanger.
o
Check the water lines to ensure they are properly connected to
the correct inlet and outlet ports
o
Check the water lines to ensure they are not kinked or blocked.
o
Try to prime the circulation system using the H2O Prime
function.
•
If using Heater Cooler:
o
Ensure the heater cooler unit is properly filled with water and
primed per manufacturer recommendations
o
Check the water lines to ensure they are properly connected to
the correct inlet and outlet ports
o
Check the water lines to ensure they are not kinked or blocked.
o
Ensure the heater-cooler is in the correct mode and able to flow
properly.
•
If Circulation system is having trouble priming due to excessive air, unlock heat
exchanger and insert one prong of the hemostat into HEX port to remove air
locked in Circulation system.
11.4.8.Cannot Close the Door
•
If the door is opened while the delivery set is primed, blood will fill the cassette
and make it difficult to close the door. To close the door, clamp the source lines,
manually open the vent valve and depress both cassette chambers until most of
the fluid is evacuated. Release the vent valve and close the door.
11.4.9.Cannot Install Heat Exchanger
•
Refer to section 8.1.1 for full details on installation of the heat exchanger.
•
Verify that the locking knob is fully unlocked by rotating it counterclockwise
(CCW).
36
Содержание MPS 3
Страница 1: ...Quest Medical Inc...
Страница 11: ...MPS 3 System Introduction TTBD Time to be Delivered Term Definition Description 5...
Страница 12: ...MPS 3 System Introduction 6...
Страница 30: ...stem System Specifications and Performance This page intentionally left blank 8...
Страница 42: ...MPS 3 System Warnings and Cautions This page intentionally left blank 12...
Страница 60: ...MPS 3 System System Installation 18...
Страница 64: ...MPS 3 System System Installation Figure 21 Tighten Pole Clamp to Secure Mounting Arm 22...
Страница 90: ...MPS 3 System Startup This page intentionally left blank 9...
Страница 123: ...MPS 3 System Graphical User Interface GUI Overview Figure 80 Auto Mode Screen 33...
Страница 135: ...MPS 3 System Disposable Installation Figure 96 Align the Additive Cartridge 12...
Страница 149: ...MPS 3 System Functional Use Figure 108 New Case Parameters Page 2 Figure 109 New Case Parameters Page 3 3...
Страница 150: ...MPS 3 System Functional Use Figure 110 New Case Parameters Page 4 Figure 111 New Case Parameters Page 5 4...
Страница 151: ...MPS 3 System Functional Use Figure 112 New Case Parameters Page 6 Figure 113 New Case Parameters Page 7 5...
Страница 152: ...MPS 3 System Functional Use Figure 114 New Case Setup Parameters Page 8 6...
Страница 167: ...MPS 3 System Functional Use Figure 130 Dose History Screen 21...
Страница 171: ...MPS 3 System Functional Use Figure 133 Case Report 1 of 4 Figure 134 Case Report Procedure View 25...
Страница 175: ...MPS 3 System Functional Use Figure 136 Case Report 2 of 4 29...
Страница 178: ...MPS 3 System Functional Use 32...
Страница 185: ...MPS 3 System Functional Use Figure 149 Error Report 39...
Страница 187: ...MPS 3 System Functional Use Figure 150 Entering Simulator Mode Warning Figure 151 Simulator Mode 41...
Страница 231: ...MPS 3 System Troubleshooting 40...
Страница 232: ...MPS 3 System Troubleshooting Figure 157 Alarm 314 Figure 158 Alarm 315 41...
Страница 233: ...MPS 3 System Troubleshooting This page intentionally left blank 42...
Страница 234: ...MPS 3 System Troubleshooting 43...
Страница 247: ...MPS 3 System System Maintenance Figure 165 Data Logs Screen 13...
Страница 252: ...MPS 3 System Service 4...
Страница 256: ...MPS 3 System Part Numbers and Accessories 4...
Страница 258: ...MPS 3 System Part Numbers and Accessories 6...
Страница 259: ...MPS 3 System Part Numbers and Accessories 7...
Страница 260: ......
Страница 268: ...MPS 3 System Appendix This page intentionally left blank 8...