TECHNICAL SPECIFICATION

Dose power indication, range of

µSv/h

0,05 to 999

Energy range of registered:

Gamma radiation

X-radiation

Beta-radiation

MeV

0,1 to 1,25
0,03 to 3,0

0,4 to 3,5

Error,

where Р – dose power in µSv/h

%

± (15+6/Р)

Alarm threshold (in increments of 0.05)

µSv/h

0,05 to 1,2

Measurement time

sec

10

Reading indication

continuous

Batteries («AAA»)

pcs

2

Uninterrupted operation time (about)

h

~ 300

Temperature range (max humidity 85% at 25C)

°С

-18 to +45

Device dimensions

mm

97x68x24

Weight (without batteries)

kg

0,08

* At factory settings, with normal background conditions, without use of 

flashlight.

Factory settings

Threshold

- 0,30 µSv/h

Audible alarm

- on

Date collection

- off

Vibration signal

- off

Backlight

- off

Device appearance

Device operation

Setup

RADEX RD1212

User Guide

9

Time and date

Setting TIME

1. Select

 menu 

 

setup

   

time and date

.

2. Move   to the first line (time) and press  .
3. Use   or   to set hours and press  .
4. Use   or   to set minutes and press  .

Setting DATE

1. Select

 menu 

 

setup

   

time and date

.

2. Move   to the second line (date) and press  .
3. Use   or   to set day and press  .
4. Use   or   to set month and press  .
5. Use   or   to set year and press  .

Language

Language Selection

1. 

menu   setup   language.

2. Move   to selected language and press  , a tick   

symbol will appear next to selected 

language

.

FLASHLIGHT

To turn the flashlight ON/OFF, press   and hold for a few 

seconds.

When batterie are removed, the Time and Date 

settings will re-set after 40 hours.

Содержание RADEX RD1212

Страница 1: ...Radiation Detector RADEX RD1212 User Guide ...

Страница 2: ...ures radioactivity stores measurement results in memory data transfer to PC self testing Time and Date functions audible alarm vibration alarm integrated flashlight multilingual Eng De Fr Rus Jp RADEX RD1212 User Guide Please note that this device uses statistical probability function to calculate radiation values Even in identical circumstances two identical devices may show varying results The r...

Страница 3: ...f RadexRead software 5 Operation of RadexRead software 5 Settings Threshold 7 Data collection 7 Audible alarm 8 Vibration alarm 8 Backlight 8 Time and Date 9 Language 9 Flashlight 9 Miscellaneous Technical specification 9 Recommendations on surveying of objects 10 Operational restrictions 10 Symbols used in this Guide selection indicator function is Selected sequence of actions note Device operati...

Страница 4: ...DEVICE APPEARANCE Device operation Setup Miscellaneous RADEX RD1212 User Guide 1 Location of sensor Battery compartment lid Flashlight Key Mini USB Display Key Key ...

Страница 5: ...tion On Off 4 Backlight 5 Key function 6 Key function 7 Measurement result 8 Key function 9 Date 10 Time 11 Background Radiation value 12 Alarm threshold 13 Battery charge indicator high low battery should be replaced 14 Unit of measurement micro Sievert per hour 15 Blinking indicated detection of a quantum 16 Indicator of the number of measurement cycles 1 cycle 10 cycles and more ...

Страница 6: ...on Menu from the Main screen press The Main Menu will be displayed use buttons and to scroll the cursor next to desired function Use button to Select move the cursor to the desired function and Select it by pressing A tick mark will Appear Disappear next to the desired function After 30 seconds of inactivity the device will revert to Main Screen Please remove the batteries if the device is not goi...

Страница 7: ...ation was measured earlier you may proceed with measuring current radiation in excess of background 1 Enter menu background 2 Move to on position and press DEVICE OPERATION Measurements Measurement begins when the device is turned On The first result will appear on display after 10 seconds Measurements with Background Function While taking measurements with Background Function two radiation values...

Страница 8: ...lect Install RadexRead 3 Follow the prompts of the installation Wizard to install RadexRead application Operation with RadexRead program RadexRead allows you to perform the following functions view measurement results stored in memory save results in TXT and CSV formats delete measurement results from the device memory Connecting the device to PC 1 Turn On the device 2 Connect the device to PC usi...

Страница 9: ...Start RadexRead program 2 Select SAVE a dialog box will open 3 Select path to save the file 4 Enter filename 5 Select a file format in which the measurement results will be stored 6 Click SAVE Deleting data from PC memory 1 Start RadexRead program 2 click DELETE Device appearance Setup Miscellaneous RADEX RD1212 User Guide 6 ...

Страница 10: ...2 User Guide 7 SETUP Threshold This function allows to set dose power level value which will trigger an alarm Audible or Vibration Turning ON Alarm function 1 Select menu levels 2 Move to on position and press Turning OFF Alarm function 1 Select menu levels 2 Move to off position and press Setting threshold level to trigger an alarm 1 Select menu levels 2 Move to a numerical value and press 3 By p...

Страница 11: ... function allows the vibration signal to respond to registration of a quantum In case a threshold function is turned ON the vibration signal is issued when registered value exceeds the threshold level Switching ON vibration alarm 1 Select menu setup 2 Move to vibra call and press a tick symbol will appear next to vibra call Switching OFF vibration alarm 1 Select menu setup 2 Move to vibra call and...

Страница 12: ...gs Threshold 0 30 µSv h Audible alarm on Date collection off Vibration signal off Backlight off Device appearance Device operation Setup RADEX RD1212 User Guide 9 Time and date Setting TIME 1 Select menu setup time and date 2 Move to the first line time and press 3 Use or to set hours and press 4 Use or to set minutes and press Setting DATE 1 Select menu setup time and date 2 Move to the second li...

Страница 13: ... the open site near the surveyed building in 5 locations 2 Perform measurements inside the building 3 If the registered radiation inside a building exceeds radiation at the open site by more than 0 2 mSv h this may indicate presence of unfavourable radiation environment Device appearance Device operation Setup RADEX RD1212 User Guide 10 OPERATIONAL RESTRICTIONS Do not subject the device to high te...

Отзывы: