Chapter 2 Assembly & Installation
Revision W
HF Series X-ray Generators - Service Manual
2-46
Quantum Medical Imaging
6. The Power/Expose Control Box (AY20-271) must be installed between the
Generator and the "Non-patient Area" in which the Q-VISION OCP and
the Operator will be positioned. The Power/Expose Control Box should
be securely mounted in close proximity to the Operator’s position for easy
access to the ON/ STANDBY Switch, and in clear view of the Q-VISION
OCP. A 30-inch expansion cable for the 2-stage Expose Switch is hard-
wired to the Power/Expose Control Box.
Figure 2-24c. Power/Expose Control Box
7. A prefabricated 50-ft. or 100 ft. cable with 6-pin modular plugs (AY51-
243-XXX), is supplied for connecting the Power/Expose Control Box to the
generator OCP Interface Board (AY40-073S). See Figure 2-24c. DO NOT
route cable through an opening in the generator containing a High Volt-
age Cable (see Figure 2-13).
8. On systems provided with a Wall Mount Kit (option QGV-WM), install the
wall mount components (VESA Mount Adapter, VESA Mount Plate, and
Wall Mount Bracket) in accordance with instructions provided with the
Wall Mount Kit.
9. Ensure wall attachment hardware is suitable for wall construction type
(each screw capable of supporting 25 lbs. pull load).
Integrated Workstation Operator Control Panel
If the generator is to be integrated with an external digital imaging worksta-
tion computer for operation as a generator control panel, the "Q-Connect"
option must be installed in the generator. This option provides an Ethernet
adapter to enable connection of the generator to the workstation personal
computer (PC). Typically, this adapter is factory installed, and is ready for use
when the unit is shipped. If the Q-Connect Ethernet adapter is needed,
6-PIN
MODULAR
PLUG
Содержание ODYSSEY HF Series
Страница 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Страница 15: ...Chapter 1 1 SPECIFICATIONS 1...
Страница 16: ...1 2...
Страница 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Страница 38: ...2 2...
Страница 93: ...Chapter 3 1 CALIBRATION 3...
Страница 94: ...3 2...
Страница 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Страница 166: ...4 2...
Страница 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Страница 186: ...5 2...
Страница 281: ...Chapter 6 1 DIAGRAMS 6...
Страница 282: ...6 2...
Страница 349: ...A 2...
Страница 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Страница 365: ...B 2...
Страница 420: ...C 1 Log4View Error Logger C Appendix...
Страница 421: ...C 2...
Страница 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Страница 429: ...D 2...