Chapter 2 Assembly & Installation
HF Series X-ray Generators - Service Manual
Revision W
Quantum Medical Imaging
2-39
Interfacing to Pausch Tomography System (When Applicable)
To interface the HF Series generator to Pausch tomo system, two lengths of
10-conductor shielded cable is required (supplied with the HF Series genera-
tor). No wire terminations are required for interconnecting the cables. Refer
to Figure 2-22 for Pausch tomo/generator interface wiring. After interfacing,
secure the tomo interface cable securely using cable ties as required.
Check that jumper JP1 is not installed on Universal AEC Board A11
(AY40-027S). If installed, remove this jumper.
IMPORTANT! Be sure to re-program the tomo
sweep times on the Pausch tomography system to
correspond with available time settings selectable
at the HF Series generator Operator Control
Panel.
Содержание ODYSSEY HF Series
Страница 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Страница 15: ...Chapter 1 1 SPECIFICATIONS 1...
Страница 16: ...1 2...
Страница 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Страница 38: ...2 2...
Страница 93: ...Chapter 3 1 CALIBRATION 3...
Страница 94: ...3 2...
Страница 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Страница 166: ...4 2...
Страница 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Страница 186: ...5 2...
Страница 281: ...Chapter 6 1 DIAGRAMS 6...
Страница 282: ...6 2...
Страница 349: ...A 2...
Страница 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Страница 365: ...B 2...
Страница 420: ...C 1 Log4View Error Logger C Appendix...
Страница 421: ...C 2...
Страница 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Страница 429: ...D 2...