Chapter 2 Assembly & Installation
HF Series X-ray Generators - Service Manual
Revision W
Quantum Medical Imaging
2-35
(Note: Perform ion chamber balance procedure in accordance with manufac-
turer’s instructions provided with chamber.)
Typically, when the HF Series x-ray generator is shipped with an AID ICX ion
chamber, the chamber jumpers are pre-set in the following configuration:
CAUTION! The ion chamber must be electrically
isolated from the bucky to ensure proper AEC sys-
tem performance.
For systems equipped with universal type AEC (AEC Board AY40-027S):
Up to four ion chambers can be used. Connect ion chambers as follows:
• Wall#1 ion chamber to A11J8
• Wall#2 ion chamber to A11J12
• Table ion chamber to A11J10
• Auxiliary ion chamber to A11J13
Note that to connect a Siemens ion chamber AEC cable to universal AEC
Board A11 requires an interface cable assembly AY51-061 for each chamber.
This cable assembly adapts the ion chamber connector to 9-pin D-Sub type
used on the AEC Board, and provides a separate 300 VDC wire for connection
to A11TB1.
Route the ion chamber cable(s) as necessary to exit through the access open-
ing in top rear side of generator chassis, and secure as required. DO NOT
route through an opening containing a High Voltage Cable (see Figure 2-13).
Interfacing to Canon Digital Radiographic Receptor (When
Applicable)
The digital radiographic receptor is interconnected to the HF Series x-ray gen-
erator via Interface Board A9, P/N AY40-034T terminal blocks A9TB7 (Wall
Stand) and A9TB3 (Table), respectively, using 4-conductor, 18AWG cable.
Refer to the digital receptor manufacturer’s installation manual for more
detailed wiring instructions.
Table 2-5: AID ICX-153 or ICX-1153 Ion Chamber Jumper Configuration
JUMPER
POSITION
FUNCTION
61135A W1
1-2 & 3-4
LOW ACTIVE RESET
61135A W2, W3, W4
1-2 & 3-4
LOW ACTIVE FIELD SELECT
61135A W5, W6
1-2 & 3-4
A-B-C = 2-1-3
61135A W7
2-3
POSITIVE OUTPUT SIGNAL
Содержание ODYSSEY HF Series
Страница 1: ...Service Manual ODYSSEY Q VISION HF Series X ray Generators Manual Part No DC30 011 Revision W...
Страница 15: ...Chapter 1 1 SPECIFICATIONS 1...
Страница 16: ...1 2...
Страница 37: ...Chapter 2 1 ASSEMBLY INSTALLATION 2...
Страница 38: ...2 2...
Страница 93: ...Chapter 3 1 CALIBRATION 3...
Страница 94: ...3 2...
Страница 165: ...Chapter 4 1 THEORY OF OPERATION 4...
Страница 166: ...4 2...
Страница 185: ...Chapter 5 1 SERVICE INSTRUCTIONS 5...
Страница 186: ...5 2...
Страница 281: ...Chapter 6 1 DIAGRAMS 6...
Страница 282: ...6 2...
Страница 349: ...A 2...
Страница 364: ...B 1 CALIBRATION Agfa DiRex B Appendix...
Страница 365: ...B 2...
Страница 420: ...C 1 Log4View Error Logger C Appendix...
Страница 421: ...C 2...
Страница 428: ...D 1 CALIBRATION Using Generator Service Tools D Appendix...
Страница 429: ...D 2...