background image

 
 
 

 

MPMS Service Note 1014-305 

Quantum Design 

MPMS Service Note 1014-305, Rev. C-1 

1

 

 

October 20, 2005 

Removing and Installing the Sample Tube 

The MPMS sample-tube assembly contains all the MPMS thermometry and the sample-space 
heaters. Hence, you will need to replace the sample tube in the event a thermometer or heater fails. 
Additionally, some MPMS options, such as the Enhanced Temperature Control (ETC) option, 
require the installation of a sample tube that has specialized components. 

In preparation for removing and reinstalling the MPMS sample tube, you will remove or disconnect 
various related components, such as the sample, sample transport, top plate, and head-cover 
assembly. However, you will leave the probe in the dewar. 

The preparatory procedures are described in Section I. Section II has the instructions for reinstalling 
the components and the sample tube, and Section III has instructions for restoring the system and 
its software. 

We strongly recommend that you gather your tools and review all the instructions before you begin 
so that you can prepare for the different steps. Your preparation is especially important for removing 
and working around parts with fragile components and for performing the replacement quickly. 

Important:

 You must complete these procedures as quickly as possible (no more than 20 min) to 

minimize the chance that an ice plug will form in the impedance. 

Tools and Supplies 

Allen wrench, sizes 1/16, 3/16, 3/32, 5/64, 7/64 

vacuum grease (Apiezon M) 

I. 

Removing the Sample Tube 

I.A  Remove the Sample 

1.  Set the system to room temperature (298 K) and the field to zero (H=0 Oe). Wait for the 

system temperature to stabilize. 

2.  Remove the sample. 

3.  Exit from MultiVu (or MPMSR2). 

4.  Disconnect the gas lines from the probe. Leave the lines leaning against the connectors. 

Содержание MPMS

Страница 1: ... the instructions for reinstalling the components and the sample tube and Section III has instructions for restoring the system and its software We strongly recommend that you gather your tools and review all the instructions before you begin so that you can prepare for the different steps Your preparation is especially important for removing and working around parts with fragile components and fo...

Страница 2: ...u will be removing the four screws E Figure 2 that fasten all the probe head assemblies together Be careful not to shift the assembly components before it is time to remove them in Section I F below 1 Loosen and remove the four 1 4 20 cap screws E Figure 2 from the top of the top plate F Figure 2 2 Loosen and remove the two 6 32 button head screws G Figure 2 from the back of the top plate 3 Remove...

Страница 3: ... on the vacuum sleeve assembly 4 Reattach the cable assembly O Figure 2 to the connector N Figure 2 on the sample tube assembly M Figure 2 Note that the brown wire connects to pin number 1 and is aligned with the arrow on the connector socket of the sample tube assembly 5 Move back the airlock valve assembly I Figure 2 until it is properly seated on the sample tube assembly II B Reinstall the Top ...

Страница 4: ...e tube has a unique set of ipl files that contain the thermometry calibration data After you restart the software and set up the system you will copy the ipl files for the new sample tube to your system The restoration procedures vary according to the software that is being used Section III A has instructions for systems that use MPMSR2 Section III B has instructions for systems that use MultiVu I...

Страница 5: ... Chamber Gas Control Diagnostics dialog locate the Valves subsection c Select Open in the options menu next to the Cooling valve d Click on the Set button Important You must click on the Set button or the Cooling valve will not be opened e Click on the Close button to close the Chamber Gas Control Diagnostics dialog Figure 1 MPMS MultiVu Chamber Gas Control Diagnostics dialog 3 Reconnect the large...

Страница 6: ...Removing and Installing the MPMS Sample Tube 6 MPMS Service Note 1014 305 Rev C 1 Quantum Design October 20 2005 Figure 1 Sample transport ...

Страница 7: ...Removing and Installing the MPMS Sample Tube Quantum Design MPMS Service Note 1014 305 Rev C 1 7 October 20 2005 Figure 2 Components of probe head ...

Отзывы: