MINISYSTEM Manual Version 1.5
Page 32
IL #,####
Illuminator level. First parameter is the illuminator channel. 0 =
Coax(if installed), 1 = Slit backlight. The second parameter is the
intensity level 0 to 1000.
Status returns level for both illuminators.
Calibration Commands
ECS #,#
ECS?
Energy Calibration Setup. Sets up the system for energy
calibration. First parameter is the wavelength; second parameter
is high/low energy mode. Output energy will automatically be set
to 50%.
ECM #
ECM?
Energy Calibration Maximum. Parameter is the measured energy
when the output energy is set to 50% level using the ECS
command.
NOTE: This will be measured at high energy mode only, and will
be done for each wavelength available on the system.
ECF #,#
ECF?
Energy Calibration Full scale. Sets the desired full scale energy
output for the selected wavelength at high level output. First
parameter is wavelength; second parameter is the desired high
level full scale energy. The low value will automatically be scaled
to 25% of the high setting. A query returns both the high and low
values for the current wavelength. High, Low.
ECR
Energy Calibration Reset. Resets values to defaults for all
wavelengths.