MINISYSTEM Manual Version 1.5
Page 30
UP #
UP?
User Password. This command can be used to query or change
the user password to allow enabling and firing of the laser. The
Master code must be entered prior to using this command. See the
PW command.
VN?
Version Number
– Query Only. Returns the current LSC version
number in the format of major.minor.release. Ex: 1.3.6
Laser Operation Commands
BC #
BC?
Burst Count. Sets the number of laser shots to be fired when
firing mode is set to burst. Parameters are from 1-1000. Default
is 10. A query will return the currently set burst count.
BL #,#
BL:PE?
BL:SE?
Blend (if installed). First parameter is blend percentage of Primary
to Total energy output amount (Primary/Total). Second parameter
is the overall energy density. Range is 0-1000 for both. A query
returns both the blend percentage and the energy density.
Command only functional on blend systems.
ED #,#,###
ED?
Set Energy Density. Parameters: First parameter selects
wavelength. Range is 1 to 4 where 1 = IR, 2 = GRN, 3 = UV, 4 =
DUV. Second parameter selects High (1) or Low (0). High mode
gives 100% of full scale and Low mode is 25% of full scale. Third
parameter is Energy Density (0
– 1000). Where 1000 equals the
wavelength/mode specific maximum output. ED? returns current
energy density setting. The format of the query is #,#,####.
If the system is a blend type, then the parameters are as follows:
First parameter selects the blend percentage. Second parameter
selects energy mode. Third parameter selects energy density.
EM #
EM?
Energy Mode. Sets the type of energy mode on systems that
support different energy modes. 0 = Normal mode, 1 = High
mode, 2 = Weld mode. If the system does not support High or
Weld mode, only Normal mode can be set.
EN #
EN?
Enable. This will enable or disable the laser. The laser must first
be enabled before firing. 0 = Disabled, 1 = Enabled. (Default: 0).
A query returns the current enable state.
ET #
ET?
External Trigger. Sets the external trigger mode. 0 = Off, 1 = On,
2 = Footswitch/Gate
FL #
FL?
Fire Laser. Starts the laser pulsing.
0 = Stop Firing Laser, place laser in idle state, shutter will be
closed.
1 = Fire laser,.
2 = Fire laser, shutter is open.
A query returns the current firing state.