Quantum Composers 9530 Series Скачать руководство пользователя страница 4

MODULE SPECIFICATIONS

 

Toll Free Phone

  

(800) 510-6530 

 

Email

  

[email protected]

 

Fax Phone Line

  

(406) 582-0237 

 

Web 

www.QuantumComposers.com

TTL/ADJUSTABLE CHANNEL OUTPUTS

oputput impedance    

50 ohm

TTL/CMOS MODES

output level    

 

4.0 V typ into 1 kohm

rise time  

 

 

3 ns typ

slew rate  

 

 

0.5 V/ns

jitter    

 

 

50 ps RMS

ADJUSTABLE MODE

output level    

 

2.0 to 20 VDC into 1 kohm

 

 

 

 

1.0 to 10 VDC into 50 ohm

output resolution  

 

10 mV

current   

 

 

200 mA typical, 400 mA max

    (short 

pulses)

slew rate  

 

 

> 0.1 V/ns

overshoot  

 

 

< 100 mV + 10 % of pulse  

 

    amplitude
rise time 

 

 

15 ns typ @ 20 V (high imp) 

 

 

 

 

25 ns typ @ 10 V (50 ohm) 

 

 

 

 

(10 % - 90 %)

TRIGGER/GATE DUAL INPUT (STANDARD)

Standard dual channel input, providing one trigger input and one 

gate input. May be used with the dual trigger firmware option to 

provide two independent trigger sources. 

threshold  

 

 

0.2 to 15 VDC

maximum input voltage  

60 V peak

resolution  

 

 

10 mV

input impedance  

 

1 Mohm + 40 pF or 50 ohm

insertion delay   

 

< 160 ns

pulse inhibit delay  

 

< 120 ns

output inhibit delay    

< 50 ns

jitter    

 

 

< 800 ps RMS

*Other custom modules available. Call with your request.

    

SYSTEM EXTERNAL TRIGGER/GATE INPUT(S)
TRIGGER INPUT

function 

 

 

generate individual pulses, start a burst  

    or 

continuous 

stream

rate  

 

 

 

DC to 1/(200 ns + longest active

    pulse)
slope    

 

 

rising or falling (maximum of 5 MHz)

behavior 

 

 

used to control the internal rate

GATE INPUT

function   

 

pulse inhibit or output inhibit

polarity    

 

active high / active low

behavior   

 

used to control the internal rate generator

STANDARD FEATURES & FUNCTIONS

communications  

 

USB/RS232/Ethernet

external clock in  

 

10 MHz - 100 MHz  in 1 MHz increments

external clock out  

 

5 MHz - 40 MHz

configuration storage   

T0, Rate, Chan, 2x ExtPLL, 1 ExtPLL, ½    

 

 

 

 

ExtPLL, ½ Ext, 40MHz, 20MHz, 10MHz,    

    5MHz, 

and 

Disabled

    

STANDARD OUTPUT MODULES 

AT20    

 

 

quad channel, TTL/CMOS &

    adjustable 

output 

module

OPTIONAL MODULE

TZ50    

 

 

quad channel, high current  

 

 

 

 

 

 

TTL/CMOS (for driving 50 ohm loads)    

    & 

adjustable 

output 

module

SYSTEM OPTIONS

 

 

 

incrementing (provides automatic  

 

    high 

speed 

incrementing/decrementing 

 

 

 

 

 

of delay and/or pulsewidth for each    

    channel)

DT15    

 

 

dual trigger logic – provides

 

 

 

 

additional trigger via gate input

  

V1.2 3/1/12

Содержание 9530 Series

Страница 1: ...des the latest in laser timing and synchronization Offering a unique 1U 19 rackmount package with all rear panel connections it is well suited for integration into your rack timing and control systems Key Features 250 ps Timing Resolution 50 ps Channel to Channel Jitter 1U Rackmount Ready Easy Programming Interface 4 or 8 Independent Channel Outputs Free LabVIEW Driver Ethernet USB RS232 Standard ...

Страница 2: ...lexibility in setting up an experiment or synchronizing multiple events Complimentary NI certified LabVIEW drivers available Advanced Features Options Clock input output allows master clock input from 10 MHz to 100 MHz with complete system timing relative to that signal with low jitter Field programmability custom features upgrades and fixes via fully programmable FPGA Settings Programming saved o...

Страница 3: ...h or shutter is enabled Single shot or Burst mode laser pulse bursts controlling either just the Q switch or entire laser Duty cycle mode allows firing laser at an optimal rate but picking pulses out at the user required rate Output multiplexer allows the timing of any combination of channels to be output on any of the output ports providing very complex pulse trains I O CONFIGURATION Models Outpu...

Страница 4: ...y 120 ns output inhibit delay 50 ns jitter 800 ps RMS Other custom modules available Call with your request SYSTEM EXTERNAL TRIGGER GATE INPUT S TRIGGER INPUT function generate individual pulses start a burst or continuous stream rate DC to 1 200 ns longest active pulse slope rising or falling maximum of 5 MHz behavior used to control the internal rate GATE INPUT function pulse inhibit or output i...

Отзывы: