11
|
Ultra
Version M 22-April-2012
|
Ultra
3.0 Filling the Standard ICE450
CAUTION:
Do not operate the system until it has been filled. Running
the pump without coolant may damage the pump. Running the pump
without the Laser Head connected may result in pump failure due to
overpressure.
Complete all preliminary steps prior to laser operation, including visual
inspection for damage, filling coolant, properly connecting all cables,
and performing all laser safety precautions.
Always drain the system before shipping or storage. See
Use only distilled water with 1M
Ω
-cm to 5M
Ω
-cm resistivity.
1. Connect the fill/drain tube and vent tube to the fill/drain bottle.
Press firmly until you hear the fitting snap into place.
2. Connect the fill/drain tube to the ICE front panel Fill/Drain fitting.
Connect the vent tube to the ICE Vent fitting.
3. Remove the cap and fill the bottle with the approved coolant.
4. Raise the bottle higher than the ICE reservoir. Keep the fill bottle
cap loose.
5. Allow coolant to flow into the ICE reservoir until it rises into the
vent tube.
6. Turn the Key Switch ON (“I”) to power the system.
The pump turns ON automatically when the system is powered ON.
Coolant from the ICE reservoir flows into the coolant lines. When the
coolant falls below the minimum level, the pump automatically turns
OFF and the reservoir-level window lights blink.
7. Using the Remote Box System menu, set the pump to Fill mode.
.
8. Refill the bottle. Continue adding coolant until the level is well
above the reservoir-level window minimum mark. When the fluid
level is sufficient, the pump automatically turns ON and again fills
the lines with coolant.
Note:
If more than three pump cycles are needed to fill the lines, turn the
power OFF. Wait five seconds, then turn the system ON again.
The reservoir-level light remains ON whenever the coolant is above the
required minimum.
9. Filling is complete when the coolant level is stable and the pump
remains ON.
Figure 4:
Filling a Standard ICE450
Vent tube
Fill/drain
bottle
ICE450
Fill tube
Vent
fitting
Fill/Drain fitting
Reservoir-level window
Key
switch
Coolant Loop
(only used when draining)
Fill/Drain
tube
Vent tube
Vent tube
10. Using the Remote Box System menu, set the pump to Normal mode. See
11. Replace the fill/drain bottle reservoir cap.
12. Disconnect the fill/drain bottle from the ICE front panel. The tubing from the
fill/drain bottle may leak. Have a container ready.
13. Purge air bubbles from the Laser Head by tilting it vertically with the
coolant circulating for 10 seconds.
Note:
Any time the Laser Head has been disconnected and reconnected, follow this
procedure to purge all of the air from the Laser Head and coolant lines before
operating the laser.
14. Remove the protective sticker from the Laser Head before operating the
system.