Warrior Extreme Blood and Fluid Warmer Instructions for Use
System Components
QIF-IFU-00001 Rev. A6
Page 7 of 28
Figure 4. Compact Disposable Unit
Intended Users
Who should use the Warrior Extreme Blood and Fluid Warmer?
The Warrior Extreme should only be used by qualified healthcare professionals that have
read the training materials and this IFU, and fully understand how to operate this system.
Fluid Warming Capacity
How much fluid can the Warrior Extreme warm using a fully charged battery?
The amount of fluid the Warrior Extreme can warm depends on the battery charge level,
inlet fluid temperature and flow rate.
When using a fully charged battery, at flow rate of 160–180 ml/min, the Warrior Extreme
can warm up to three liters of parenterally administrated fluid with an inlet temperature
of 4 °C (39.2 °F).
When there are extreme changes in the flow rate or the inlet fluid temperature, the
Warrior Extreme will require an adjustment period of up to 20 seconds. A set of internal
2
1
3
3
2
1
1
Inlet Luer
2
Outlet Luer
3
Connector to Base Unit or Extension Cable
Connecting
Cable
Out
In
Base Unit
Figure 5. Base Unit connected to the Compact
Disposable Unit