Mini 10K Centrifuge
Страница 1: ...Qor Labs www QorLabs com 678 218 4552 Mini 10K Centrifuge ...
Страница 2: ...he centrifuge Place the centrifuge on a sturdy level surface Use the plastic latch to open the lid Verify that there are no loose objects or packing material in the tube chamber DO NOT LOAD TUBES AT THIS TIME Verify that the rotor is secure by pressing down firmly Verify that there are no tubes or inserts in the rotor Close the lid ensuring that it clicks and locks Verify that the ON OFF switch is...
Страница 3: ...ight are placed opposite each other Use a water filled tube as a balance tube if necessary Proper balancing will improve sample separation and extend the life of the centrifuge Spinning out of balance loads may break tubes or damage the centrifuge Micro tubes and PCR strips are intended to hang by their caps in the centrifuge rotor Be sure to use good quality tubes that will withstand forces up to...
Страница 4: ...es in a centrifuge it is recommended that rotors be inspected monthly for wear and fatigue If there is any indication of wear the rotor should be removed from service Contact Qor Labs for return instructions so the rotor can be evaluated by a technician for repair or replacement After 2 years of service it is recommended that rotors be returned to Qor Labs for inspection or replacement Speed Range...