Operating Procedures
PyroMark Q96 ID User Manual 01/2016
5-15
5.
Pipet the recommended volume of enzyme solution into
the
E
compartment
6.
Pipet the recommended volume of each dNTP solution
into the
A
,
C
,
G
, and
T
compartments.
Note
: do not use filtered tips to pipet nucleotides.
PyroMark Q96 Cartridge (from above) with indicated
compartments for enzymes, substrate, and nucleotides
.
5.5
Run
5.5.1
Loading the PyroMark Q96 Plate Low
1.
Open the instrument lid.
Note
: The dispensing unit and process chamber will
move if a run is accidentally started.
2.
Open the process chamber lid.
3.
Open the plate-holding frame.
4.
Place the PyroMark Q96 Plate Low on the heating block.
5.
Close the plate-holding frame and the process chamber.
Label
A
G
T
C
E
S
Содержание PyroMark Q96 ID
Страница 1: ...Version 5 January 2016 Sample Assay Technologies PyroMark Q96 ID User Manual...
Страница 2: ......
Страница 8: ...Contents Contents 6 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 14: ...Safety Information 1 6 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 32: ...Installation Procedures 4 6 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 58: ...Operating Procedures 5 26 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 78: ...Maintenance 7 12 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 96: ...Appendix A A 8 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 106: ...Appendix C C 2 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 114: ...Appendix D D 8 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 122: ...Appendix E E 8 PyroMark Q96 ID User Manual 01 2016 This page intentionally left blank...
Страница 125: ......
Страница 126: ...1100446 01 2016 Sample Assay Technologies...