EZ1 Advanced XL User Manual 11/2017
44
10.
Select
Options
and open the EZ1 Advanced Serial Port window .
11.
Depending on how many instruments you are using, select COM4–COM7 starting with
COM4. Depending on the PC you are using, COM3–COM6 may appear.
The EZ1 Advanced Communicator software and the PC Windows software on the PC are now
configured for use with multiple EZ1 Advanced XL and/or EZ1 Advanced instruments.
In order to check the connectivity between the PC and EZ1 Advanced XL and/or EZ1 Advanced
instruments, use the test described in Section 5.9.3.
4.6
Transportation and removal of the EZ1 Advanced XL
If the EZ1 Advanced XL needs to be transported, please follow your organization’s instructions and
guidelines. In addition, follow the steps below.
1.
Decontaminate the instrument.
2.
Pack the EZ1 Advanced XL by following the steps of the unpacking procedure described in
Section 4.3.1 in reverse order. Use original packaging material.
Appendix A gives information about transportation conditions.
If the EZ1 Advanced XL needs to be sent to QIAGEN, please contact your local instrument service
organization. Follow the steps below.
1.
Decontaminate the instrument.
2.
Follow the instruction of your local instrument service organization.
WARNING
Risk of personal injury
Two people must carry the instrument. Use the handle attached to the box to lift
the EZ1 Advanced XL. After the EZ1 Advanced XL has been unpacked, 2 people
must lift the instrument. Lift the instrument by placing your hands underneath the
bottom of the instrument.