EZ1 Advanced XL User Manual 11/2017
36
4.3.3
Configuring and setting up the EZ1 Advanced XL
Setting the installation date
The first time the EZ1 Advanced XL is switched on, it is necessary to enter the installation date. This
date will be recorded in the report file described below. The installation date is also used to let the
instrument know when to show the weekly and annual maintenance reminders.
SERV: SETUP SYSTEM
Installation date
DD MM YYYY
Key: Up, Dn, SHIFT, ENT
1.
Set the day, month and year.
2.
Press
SHIFT-down arrow
to move the cursor to the right, from DD (day) to MM (month) to
YYYY (year).
3.
Press
SHIFT-up arrow
to move the cursor left, from YYYY to MM to DD.
4.
Press the
up arrow
or
down arrow
to increase or decrease the value of the selected field.
5.
After setting the date, store the setting by pressing
ENT
.
If you accidentally set the wrong date, call QIAGEN Technical Services for assistance.
Setting the serial port
Set the serial port accordingly for use with the dot matrix printer or a PC as the output device for
the report file (Section 5.10.3, page 68).
Setting time and date
The EZ1 Advanced XL has an integrated clock and date device. Check the time and date and reset
if necessary (Section 5.10.1, page 67 and Section 5.10.2, page 67).
Setting the annual maintenance reminder
The EZ1 Advanced XL reminds you when the annual maintenance is due. You can set the reminder
to remind you annually or semiannually (Section 5.10.4, page 68).