EZ1 Advanced XL User Manual 11/2017
30
4.2.3
Installation of AC power cord
Connect one end of the AC power cord to the socket located on the rear of the EZ1 Advanced XL
and the other end to the AC power outlet.
Note
: Do not use items other than attached accessories.
The fuse box of the EZ1 Advanced XL is located under the socket for the power cord and contains
2 fuses, which are labeled as follows:
110–120 V
A 6.3 A (250 V) time-delay fuse compatible with 100–120 V power supplies
220–240 V
A 3.15 A (250 V) time-delay fuse compatible with 200–240 V power supplies
Before connecting the instrument to the power supply, you may need to rotate the fuse box in order
to select the correct fuse. A fuse is correctly selected when its label is readable and closer to the
bottom of the instrument.
Selecting the correct fuse for Selecting the correct fuse for
200–240 V-line power supply. 100–120 V line power supply.
WARNING
Electrical hazard
Never install a fuse different from that specified in the user manual.