Operation Manual
GEM and GEM PRO
OMiQGEM-L Issue.02
Page 33 of 51
Note: In the event of malfunction or damage, please contact QED or an authorized representative.
Note: For further information please contact Technical Support at QED on:
WATER FILTERS
Each inlet port has a PTFE filter to stop liquid from entering the instrument:
1)
Using the supplied tool, select the Phillips No. 1 screwdriver and unscrew the 4 faceplate screws
2)
Gently lift the faceplate away from the case, whilst keeping it parallel to the top of the case
Note: Filters firmly located in the i-Q body can be removed using the supplied filter removal tool.
3)
Replace the filters by placing them inside the faceplate.
4)
To reassemble, follow the above steps in reverse order, paying attention to the good fit of mating
components and taking care not to overtighten any screws.
QED Environmental Systems Ltd.
+44(0)333 800 0088
QED Environmental Systems, Inc.
+1 (800) 624-2026
iQV14
Содержание LANDTEC i-Q Series
Страница 1: ...Operation Manual GEM and GEM PRO...
Страница 12: ...Operation Manual GEM and GEM PRO OMiQGEM L Issue 02 Page 12 of 51 GEM VS GEM PRO...
Страница 21: ...Operation Manual GEM and GEM PRO OMiQGEM L Issue 02 Page 21 of 51 iQV07 iQV08 iQV09...
Страница 22: ...Operation Manual GEM and GEM PRO OMiQGEM L Issue 02 Page 22 of 51 iQV10...
Страница 28: ...Operation Manual GEM and GEM PRO OMiQGEM L Issue 02 Page 28 of 51 Figure 7 i Q Setup...
Страница 30: ...Operation Manual GEM and GEM PRO OMiQGEM L Issue 02 Page 30 of 51 Flow Chart 2...
Страница 48: ...Operation Manual GEM and GEM PRO OMiQGEM L Issue 02 Page 48 of 51 WARRANTY INFORMATION...
Страница 49: ...Operation Manual GEM and GEM PRO OMiQGEM L Issue 02 Page 49 of 51...