Q-SENSE AB 2007
Q-SENSE E1 LIQUID HANDLING SET UP
Q-Sense E1 Pump and Flow System Setup
Part 1 – Ismatec Reglo
Analog
Before switching the pump
on, check the voltage
setting and fuse on the
backside:
Dec 2007, Rev A
220-240V: use 250mA fuses
110-120V: use 500mA fuses
1
Insert the 3-stop color-coded tubing into the cassettes.
Voltage settings
window
2
Set the pump speed by
adjusting the 2-digit selector.
The pump tubing supplied by
Q-Sense has an ID of 0.38
mm, giving the following flow
rates:
3.2 rpm = 28 µl/min (select “01”)
160 rpm = 1.4 ml/min (select “99”)
For details and calibration function, see supplier’s manual.
Part 2 – Tubing connectors
The ferrule forms a tight
hen being screwed into
e
ed
d in
dicate the
direction
of
le
on.
seal as it is deformed
w
the flow module. Place th
ferrule and threaded
connector as shown
before attaching them.
Make sure to cut the
PTFE tubing at right
angles. It is recommend
to use the blue tube
cutting device include
the Liquid Handling Set.
Note
that the arrows on
the top of the module
in
the flow for correct samp
temperature equilibrati
3
4
Out
In