Q-SENSE AB 2007 

 

Q-SENSE E1 LIQUID HANDLING SET UP

Q-Sense E1 Pump and Flow System Setup 

 
Part 1 – Ismatec Reglo 

Analog

 

 

Before switching the pump 
on, check the voltage 
setting and fuse on the 
backside: 

 

 

 

Dec 2007, Rev A

 

 

 

 

 

 

220-240V: use 250mA fuses 
110-120V: use 500mA fuses 

                

                

 

    

 

 

 

 
Insert the 3-stop color-coded tubing into the cassettes.  

 

 

Voltage settings 

window

 

 

Set the pump speed by 
adjusting the 2-digit selector. 
The pump tubing supplied by 
Q-Sense has an ID of 0.38 
mm, giving the following flow 
rates: 

 

3.2 rpm = 28 µl/min (select “01”) 
160 rpm = 1.4 ml/min (select “99”) 

 

 
 
 
  

 

 

For details and calibration function, see supplier’s manual. 

 

 

Part 2 – Tubing connectors 

 

The ferrule forms a tight 

hen being screwed into 

ed 

d in 

dicate the 

direction

 of 

le 

on. 

                  

seal as it is deformed 
w
the flow module. Place th
ferrule and threaded 
connector as shown 
before attaching them. 
Make sure to cut the 
PTFE tubing at right 
angles. It is recommend
to use the blue tube 
cutting device include
the Liquid Handling Set. 
 

Note

 that the arrows on 

the top of the module 
in
the flow for correct samp
temperature equilibrati

 

Out 

In 