5.3
Calibration Screen and Reading
When connected, the
uPyxis®
Mobile App will default to the
Calibration
screen. From the
Calibration
screen, you can perform calibrations by pressing on
Zero Calibration
,
Air Calibration
, and
4–20mA Span
.
*
NOTE
*
Before calibrating, remove the ST-773 sensor from the water and wipe it with a
damp cloth to remove debris and any biofouling. If there is water on the membrane, dry it
with a soft cloth, taking special care not to damage the membrane.
Figure 8.
5.3.1
Zero Calibration
1. Place the sensor in Nitrogen Gas or 5% Sodium Sulfite solution (anaerobic water).
2. Let the ST-773 sensor stand for a minimum of 15 minutes. Observe the values being displayed on the
Calibration
screen.
3. Once the displayed oxygen and temperature values are stable, press
Zero Calibration
to perform a
zero calibration.
4. If the calibration is successful, the interface will return a message ”Calibration Succeeded”. If the
calibration fails, press
Zero Calibration
again and repeat.
ST-773 User Manual
| +1 (866) 203-8397
13
Содержание 53709
Страница 1: ......