Figure 8.
Typical wiring for
Advantage Controls MegaTronXS
*
NOTE
*
To get a Pitting signal, a jumper wire is needed from 4–20mA Input 3- to power
-24V.
4.2
Connecting via Bluetooth
A Bluetooth adapter (P/N: MA-CR) can be used to connect the CR-301sensor to a smart phone with the
uPyxis®
Mobile App or a computer with the
uPyxis®
Desktop App. The power should be sourced from a 24
VDC power terminal of a controller. If a controller is not available, the user may also use Pyxis PowerPACK-1
or PowerPACK-4 as an alternative to both an external power supply and a Bluetooth adapter. PowerPACK
from Pyxis Lab offers external power, Input/Output signal, and Bluetooth connectivity.
Figure 9.
Bluetooth connection to CR-301 sensor
CR-301 User Manual
| +1 (866) 203-8397