after the successful installation a new program group "Pyro
Oxygen Logger" is added to the start menu, and a short-
cut named "Oxygen Logger" can be found on the desktop
•
connect one end of the USB-interface cable
FSO2-MINI-
USB
to the connector X1 of the
FireStingO2-Mini-ST
•
connect the other end to an USB-port of the PC. The
optical port of the
FireStingO2-Mini-ST
should flash
shortly, indicating the correct startup of the
FireStingO2-
Mini-ST
•
start the
Pyro Oxygen Logger
software.
6.2
Using the Pyro Oxygen Logger software
IMPORTANT: The following instructions are only valid for the
versions ≥3.200 of the
Pyro Oxygen Logger
software. If you have
an older version, please download and install the newest version.
Please refer to the manual provided with the software for general
operation instructions.
In combination with OEM-applications it is important to know, that
the actual logger settings and the calibration data are
automatically saved in the internal flash memory of the
FireSting
device, each time the logger software is closed. A typical scenario
for OEM-applications would be, that the USB-interface cable
FSO2-MINI-USB
in combination with the
Pyro Oxygen Logger
software is used for configuring all settings and to calibrate the
connected oxygen sensor. This is done in the settings and the
calibration window, respectively.
If also the auto-mode should be configured, press within the
settings window the button Prepare Auto-Mode… (simple
settings, only for "interval auto-mode") or Advanced Auto-