USER GUIDE:
Ionic® Purification System
Purigen Biosystems
FOR RESEARCH USE ONLY.
Not for use in diagnostic procedures.
49
1.
Press the
Instrument Stats
button to view instrument statistics.
FIGURE 45:
Instrument Statistics screen
2.
Press the button on the left side to return to the
Maintenance & Service
screen.
Save Diagnostics
The
Save Diagnostics
Maintenance screen is used to save instrument diagnostics to assist in troubleshooting. The saved log file
should be sent to Purigen Biosystems. Please contact Purigen support to coordinate appropriate transfer of the log file.
If an error occurs or, if directed by Purigen Biosystems Support personnel, follow the steps below to save the instrument diagnostic log
file.
1.
From the
Maintenance & Service
menu, press
Save Diagnostics
to save the instrument diagnostic log file to a USB drive.
IMPORTANT
USB drive must be FAT32 formatted with at least 1 GB of free space.