FIGURE 36:
Clean screen -
Step 8
18.
Before removing the chip from the instrument, use a pipette to remove any remaining solution in the reservoir and cover the
reservoir with the seal provided for disposal.
19.
Dispose of the chip, reservoir, and pipettes according to your lab safety guidelines.
20.
Press the arrow button on the right side to return to the
Run Setup
screen.
Decontaminating the Ionic Purification System
If instrument contamination occurs, follow the guidelines below to decontaminate the instrument.
Interior
Ionic Purification System tray has been designed to catch and contain drips and small volume of liquid spills. Occasionally, use a soft lab
towel to clean these areas with a 70% ethanol and 30% deionized water solution. When wiping down the instrument, apply the cleaning
solution directly to a soft lab towel enough to slightly dampen it. Do not oversaturate the lab towel. Never add liquids directly onto any
surface of the instrument.
NOTE
Do not use detergents, bleach, or acetone and other harsh solvents. Apply all standard safety practices when using cleaners
and dispose of any generated waste in a responsible manner.