Altec AIR 

 

P8400W Air-dryer User’s Guide 

 

 

Page 10 of 110  

P017585 – Rev. A 

5.4 Dryer Function Overview 

 

 

Component 

Description 

Compressor 

Compresses drawn in ambient air. 

Precooler 

Cools compressed air prior to drying function. 

Unloader Valve 

Relieves excess compressor head pressure. 

Heatless Dryer 

Removes moisture from compressed air. 

Humidity Sensor 

Measures the humidity of the compressed air. 

Capacity Control Valve 

Regulates system pressure and prevents air from 
bleeding back through the heatless dryer. 

Air Tank 

Stores dry compressed air. 

Static Pressure Regulator  Regulates the static pressure (17 PSI). 

Maintains constant pressure on the flow block for 
accurate flow measuring.   

Flow Block 

Measures the flow of compressed air. 

10  Outlet Pressure Regulator  Regulates the outlet pressure. 

11  Pressure Outlet 

Outputs the pressure set by the Outlet Pressure 
Regulator. 

 

Содержание Altec AIR P8400W

Страница 1: ...P017585 Rev A P8400W Air dryer User s Guide...

Страница 2: ......

Страница 3: ...t us on the web at www AltecAIR com 2 Introduction PLEASE READ THIS USER S GUIDE THOROUGHLY AND SAVE FOR FUTURE REFERENCE This User s Guide is provided for the benefit of our customers and contains in...

Страница 4: ...tting the System Pressure 54 8 13 Setting the Static Pressure 55 8 14 Setting the Outlet Pressure 55 8 15 Engaging the Boost Transformer 56 8 16 Connecting via Web Browser 56 8 17 Using the Status Scr...

Страница 5: ...w Outlet Pressure Alarm 97 13 8 Can t Create a Low Pressure Alarm 98 13 9 High Flow Rate Alarm 98 13 10 High Cabinet Temperature Alarm 98 13 11 High Humidity 99 13 12 Can t Create a High Humidity Alar...

Страница 6: ...attention for these warnings and read them as you encounter them WARNING For your safety all the information in this User s Guide must be followed to minimize the risk of electrical shock and prevent...

Страница 7: ...g routine maintenance as outlined in the Maintaining Your Dryer section will ensure optimal performance over the lifecycle of your air dryer CAUTION Incoming power to dryer must be 190 230 VAC 1 Phase...

Страница 8: ...dependently interchangeably or simultaneously depending on your pressurized air requirement Other features of the P8400W Air dryer include a fully digital operating platform offering the most accurate...

Страница 9: ...imum 8 400 SCFD both systems 4 200 SCFD single system Power Requirements 190 230 VAC 1 Phase 50 60 Hz Electrical Characteristics Running Amps 15 20 Amp service recommended Outlet Pressure Range 0 15 P...

Страница 10: ...Humidity Sensor Measures the humidity of the compressed air 6 Capacity Control Valve Regulates system pressure and prevents air from bleeding back through the heatless dryer 7 Air Tank Stores dry com...

Страница 11: ...jewelry before performing any procedures WARNING Internal surfaces may be hot Use care when coming into contact with internal components as there is a potential for some of these components to become...

Страница 12: ...entilated and free from abrasive dust or chemicals 6 2 3 2 Ambient temperature is between 40 and 85 F optimal NOTE Higher temperatures will decrease component lifespan 6 2 3 3 Meets the following powe...

Страница 13: ...terials Large adjustable wrench Medium adjustable wrench 7 16 wrench Band cutters or snips Lifting Straps Pipe dope or pipe thread tape Cup of soapy water 1 inch paint brush recommended 6 5 Installati...

Страница 14: ...latches if a higher level of security is required 6 5 4 Check for loose parts hoses or wiring NOTE If ANY SHIPPING DAMAGE is detected file a claim with the shipping company prior to continuing the ins...

Страница 15: ...ify that the red orifice plug is still installed where shown 6 5 8 Install the two 2 purge mufflers shipped loose 6 5 9 Verify that the dryer is powered OFF 6 5 10 Plug the power cord into a 190 230 V...

Страница 16: ...measure the voltage by placing the probes between the Circuit Breaker and terminal insulation so that they touch the metal contacts If the voltage measures between 210 230 VAC skip to step 6 5 14 If...

Страница 17: ...2 Locate the Main Power Lead quick disconnects in the top section of the dryer 6 5 13 3 Unplug the Main Power Lead quick disconnects 6 5 13 4 Plug both Boost Transformer Lead connectors into the Main...

Страница 18: ...the dryer power ON NOTE Both system compressors and heatless dryers will start creating air flow through the red outlet pressure orifice 6 5 16 Set the System Pressures System 1 System 2 are adjusted...

Страница 19: ...he knob until the reading on the pressure gauge is 50 PSI 6 5 16 3 Push the knob in to lock 6 5 17 Let the dryer run until the Humidity drops under 2 on both System 1 and System 2 may take up to 15 mi...

Страница 20: ...lug System may be pressurized ALTEC AIR recommends using Installation Kit P011752 to connect your dryer to the air supply line See section 11 6 for detail 6 5 20 Turn the dryer ON 6 5 21 Set the Stati...

Страница 21: ...ssure Regulator knob out 6 5 22 3 Turn knob until Outlet Pressure OUTP reading is at the desired setting 6 5 22 4 Push knob in to lock 6 5 23 Check for air leaks NOTE This is a general procedure that...

Страница 22: ...ing or hose connection to be tested If air bubbles appear at the connection this indicates that air is leaking from the connection If any leaks are detected take steps to seal them off as necessary Ti...

Страница 23: ...o Well ventilated o Free from abrasive dust or chemicals o Ambient temperature is between 40 and 85 F optimal o Power outlet easily accessible Operating Power is between 210 230 VAC Shipping block rem...

Страница 24: ...he Limited Warranty on your product Once you register you are eligible to receive free technical support as well as updates concerning your ALTEC AIR products Register Online at www AltecAIR com regis...

Страница 25: ...is highly recommended that you remove all jewelry before performing any procedures WARNING Internal surfaces may be hot Use care when coming into contact with internal components as there is a potent...

Страница 26: ...plugs System may be pressurized ALTEC AIR recommends using Installation Kit P011752 to connect your air dryer to the air supply line See section 11 6 for detail 8 3 Powering the Dryer ON OFF 8 3 1 POW...

Страница 27: ...4 2 CENTER Button Freezes the current information screen on the display When pressed again it will allow the information screens to begin cycling again 8 4 3 Arrow Buttons Used to access navigate and...

Страница 28: ...o manually cycle through the screens by pressing the left or right arrow buttons 8 4 4 1 Tank Screen TANK Air Tank pressure fluctuates between 20 50 PSI OUTLET Outlet Pressure regulated by the Outlet...

Страница 29: ...System DUTY CYCLE The percentage of time that the compressor was on during the last compressor pump up cycle 8 4 4 4 System Runtime Screens LAST Last compressor runtime in minutes TOTAL Total compres...

Страница 30: ...et Pressure OUTP rises above the alarm set point for more than one 1 minute Default setting is 12 0 PSI Press the Info Button to see the detail of the alarm See section 13 5 for troubleshooting inform...

Страница 31: ...of the alarm See section 13 9 for troubleshooting information 8 5 4 High Compressor Temperature Alarm Occurs when the temperature in the lower compressor compartment rises above 140 F for more than on...

Страница 32: ...damage due to overheating Press the Info Button to see the detail of the alarm See section 13 10 for troubleshooting information 8 5 6 High Humidity Alarm Occurs when the Humidity level rises above th...

Страница 33: ...e alarm Default setting is 3 00 minutes Press the Info Button to see the detail of the alarm See section 13 17 for troubleshooting information 8 5 8 High Duty Cycle Alarm Occurs when the compressor is...

Страница 34: ...stem Setup Used to set system parameters including Cycle Mode Cycle Time Startup Delay Alarm Delay Date Time and Units This menu also allows for firmware updates via USB Alarm Setup Used to set the al...

Страница 35: ...to Escape to the setup screen 8 7 Using the System Setup Menu In the Setup Menu 8 7 1 Press the Up Down Arrow Buttons to Select the S in System Setup 8 7 2 Press the Enter Button to access System Setu...

Страница 36: ...he Up and Down Arrow Buttons to Change the value of the selected digit 8 7 4 5 Press the Enter Button to confirm the selected choice 8 7 4 6 Press the Left Right Arrow Buttons to Select the correct co...

Страница 37: ...fault setting is 0 seconds 8 7 6 1 Press the Enter Button to access the edit screen 8 7 6 2 Press the Left Right Arrow Buttons to Select the digit to change 8 7 6 3 Press the Up Down Arrow Buttons to...

Страница 38: ...Press the Enter Button to confirm the selected choice This will reset the Total Time to zero 0 8 7 7 4 The process is similar for resetting System 2 runtime 8 7 8 Reset Factory Default Values 8 7 8 1...

Страница 39: ...row Buttons to Select the digit to change 8 7 9 3 Press the Up Down Arrow Buttons to Change the value of the selected digit 8 7 9 4 Press the Enter Button to submit the new setting 8 7 9 5 Press the L...

Страница 40: ...row Buttons to Select the digit to change 8 7 10 3 Press the Up Down Arrow Buttons to Change the value of the selected digit 8 7 10 4 Press the Enter Button to submit the new setting 8 7 10 5 Press th...

Страница 41: ...e keyword and press the Enter Button to access the firmware update screen 8 7 11 4 Select the correct file version using the Up and Down Buttons 8 7 11 5 Press the Enter button to select the file 8 7...

Страница 42: ...new setting 8 8 Using the Alarm Setup Menu In the Setup Menu 8 8 1 Press the Up Down Arrow Buttons to Select the A in Alarm Setup 8 8 2 Press the Enter Button to access Alarm Setup 8 8 3 Set High Hum...

Страница 43: ...cted choice This will lock in the new setting 8 8 4 Set High Outlet Threshold default setting is 20 00 PSI 8 8 4 1 Press the Enter Button to access the edit screen 8 8 4 2 Press the Left Right Arrow B...

Страница 44: ...nter Button to access the edit screen 8 8 5 2 Press the Left Right Arrow Buttons to Select the digit to change 8 8 5 3 Press the Up Down Arrow Buttons to Change the value of the selected digit 8 8 5 4...

Страница 45: ...ss the Left Right Arrow Buttons to Select the digit to change 8 8 6 3 Press the Up Down Arrow Buttons to Change the value of the selected digit 8 8 6 4 Press the Enter Button when to submit the new se...

Страница 46: ...Press the Left Right Arrow Buttons to Select the digit to change 8 8 7 3 Press the Up Down Arrow Buttons to Change the value of the selected digit 8 8 7 4 Press the Enter Button when to submit the ne...

Страница 47: ...the Left Right Arrow Buttons to Select the digit to change 8 8 8 3 Press the Up Down Arrow Buttons to Change the value of the selected digit 8 8 8 4 Press the Enter Button when to submit the new setti...

Страница 48: ...k Setup 8 9 2 Enter Keyword default Keyword is 123456 8 9 2 1 Press the Left Right Arrow Buttons to Select the digit to change 8 9 2 2 Press the Up Down Arrow Buttons to Change the value of the select...

Страница 49: ...e Yes or No 8 9 3 6 Press the Enter Button to confirm the selected choice This will lock in the new setting 8 9 4 Set Subnet Mask default is 255 255 255 000 8 9 4 1 Press the Enter Button to access th...

Страница 50: ...Press the Enter Button to access the edit screen 8 9 5 2 Press the Left Right Arrow Buttons to Select the digit to change 8 9 5 3 Press the Up Down Arrow Buttons to Change the value of the selected di...

Страница 51: ...Arrow Buttons to Change the value of the selected digit 8 9 6 4 Press the Enter Button when to submit the new setting 8 9 6 5 Press the Left Right Arrow Buttons to Select the correct confirmation cho...

Страница 52: ...7 5 Press the Left Right Arrow Buttons to Select the correct confirmation choice Yes or No 8 9 7 6 Press the Enter Button to confirm the selected choice This will lock in the new settings 8 9 8 Set M...

Страница 53: ...etting 8 9 9 View MAC address 8 9 9 1 The device MAC address can be viewed from the network setup menu 8 10 Opening Panels 8 10 1 Removing Top Cover Depress the latches and pull the Top Cover off The...

Страница 54: ...vent pressure from building back up power the dryer OFF See section 8 3 for detail 8 11 4 Reinstall Front Panel 8 12 Setting the System Pressure System 1 System 2 are adjusted and set independently Wi...

Страница 55: ...Static Pressure Regulator knob out 8 13 3 Turn knob until the reading on the pressure gauge is 17 PSI 8 13 4 Push knob in to lock 8 13 5 Reinstall Front Panel 8 14 Setting the Outlet Pressure 8 14 1 R...

Страница 56: ...Boost Transformer 8 15 1 See section 6 5 13 8 16 Connecting via Web Browser Warning The intra building port Network LAN Port of the equipment is suitable for connection to intra building or unexposed...

Страница 57: ...ctly into a Laptop PC and the other end plugged into the Network Port of the air dryer Configure the network card on the Laptop PC to use the IP Address 192 168 1 102 This will make the Laptop PC comp...

Страница 58: ...larm Screen Displays all the Alarms registered by the P8400W Series Air dryer This screen is informational only Firmware Screen Allows the user to upload any software updates or upgrades to the Air dr...

Страница 59: ...the P8400W Series Air dryer Provides remote ALARM RESET Readings are displayed in GREEN unless an alarm is present Alarms are displayed in RED next to the parameter in alarm Alarm Status will display...

Страница 60: ...et Compressor Total Run Time displayed on Status Screen 8 18 Using the Setup Screen All configurations for the System Alarms Network and Keyword can be made in this screen Values in BLUE represent the...

Страница 61: ...ng a Threshold or Setup value 8 18 1 1 Click on the value to change 8 18 1 2 Type in the new value 8 18 1 3 Click the ENTER Button when done 8 18 1 4 Enter Keyword 8 18 2 Changing the Keyword 8 18 2 1...

Страница 62: ...rm Screen Displays all events such as alarms changes made and alarm resets registered by the P8400W Series Air dryer This screen is informational only You may download a csv file with all event alarm...

Страница 63: ...of the P8400W Series Air dryer Current Version Displays the current firmware version of the P8400W Series Air dryer New Version File Displays the new location and new firmware version chosen The BROW...

Страница 64: ...specific SNMP Management software used on your network This software requires a SNMP Definition Configuration File MIB file in order to properly communicate with the Air dryer The files for the P8400...

Страница 65: ...be running creating a situation for potential electrical shock It is highly recommended that you remove all jewelry before performing any procedures WARNING Internal surfaces may be hot Use care when...

Страница 66: ...e 1 System 2 Wire 8 9 2 3 Use an Amp Meter to measure the running amps With the compressor running the running amps should measure 4 0 amps or below 9 2 4 Reinstall Front Panel If the compressor measu...

Страница 67: ...be measured 9 3 3 Use a Voltmeter to measure the voltage between the Hot and Neutral wires by placing the probes in the openings in the power connector System 1 Wires 1 2 System 2 Wires 8 7 The volta...

Страница 68: ...hat they touch the metal contacts The voltage should measure 210 230 VAC 9 4 4 Reinstall Top Cover If the incoming voltage measures less than 210 VAC it is necessary to engage the Boost Transformer in...

Страница 69: ...e across the AC terminals Should measure 0 VAC 9 5 4 Use a Voltmeter to measure across the DC terminals Should measure 12 VDC With the System Compressor NOT running 9 5 5 Use a Voltmeter to measure ac...

Страница 70: ...e purge tubes from the System Heatless Dryer 9 6 3 Place your hand beneath the purge fittings to feel for purging air Air should Purge from Tower 1 side Purge from Tower 2 side 30 Seconds later Purge...

Страница 71: ...lve 9 7 3 Place your hand beneath the Unloader Valve fitting to feel for air flow Air should NOT flow from this fitting continuously Air should only be released in a short burst when the System compre...

Страница 72: ...ls from left to right 106VDC Left solenoid IN Incoming power 106VDC Right solenoid 9 8 3 Use a Voltmeter to measure the DC voltage across each set of 106VDC terminals Continue to measure for up to 45...

Страница 73: ...the Wiring Diagram section 14 1 Replace defective fan see sections 11 2 for part detail and 11 7 for ordering information 9 10 Testing Safety Relief Valve 9 10 1 Remove Front Panel see section 8 10 2...

Страница 74: ...re TANK reaches 50 0 PSI If the tank pressure TANK fails to reach 50 PSI see section 13 14 for troubleshooting information With Compressor s NOT running 9 11 4 Pull the ring handle on the Safety Relie...

Страница 75: ...2 9 12 2 Start timing when the compressor s turns on 9 12 3 Pull the ring handle on the Safety Relief Valve when necessary to keep the Tank Pressure TANK from reaching 50 PSI This prevents the compres...

Страница 76: ...n BOTH Cycle Mode see section for details on changing Cycle Mode Test one System at a time System System 1 or System 2 9 13 1 Remove Front Panel see section 8 10 2 9 13 2 Turn the System Circuit Break...

Страница 77: ...7 After three 3 minutes verify that a Humidity Alarm appears and System goes into SHUTDOWN mode 9 13 8 Replace the Humitter into the Humidity Block 9 13 9 Reinstall Front Panel 9 13 10 Turn the Syste...

Страница 78: ...of the current Outlet Pressure OUTP reading With Compressor s running 9 14 4 Pull the Outlet Pressure Regulator knob out 9 14 5 Turn knob clockwise until Outlet Pressure OUTP reading climbs over 12 0...

Страница 79: ...8 4 4 1 appears on the display press the HOLD Button on the Front Panel to freeze that screen 9 15 3 Make a note of the current Outlet Pressure OUTP reading With Compressor s running 9 15 4 Pull the O...

Страница 80: ...0 of 110 P017585 Rev A 9 15 7 Push knob in to lock 9 15 8 Press the RESET Button to clear the alarm 9 15 9 Reinstall Front Panel If you are unable to create a Low Outlet Pressure Alarm as described se...

Страница 81: ...R With Compressor s NOT running 9 16 1 Listen for any hissing sounds which may indicate a fitting or hose air leak With Compressor s running 9 16 2 Use a 1 inch paint brush to dab soapy water on the a...

Страница 82: ...n about the specific procedure Please refer to these sections for detailed procedural information NOTE When operating at higher ambient temperatures it is recommended that maintenance be performed mor...

Страница 83: ...check for any loose or disconnected wiring CAUTION SHUT DOWN IMMEDIATELY FOR REPAIRS if the air compressor s shows any evidence of overheating or presents excessive noise CAUTION Depressurizing the a...

Страница 84: ...ge must be 210 230 VAC 9 4 Test High Low Outlet Pressure Alarms 9 14 9 15 Set System Pressure 50 PSI 8 12 Set Static Pressure 17 PSI 8 13 Set Outlet Pressure 8 14 Test Consistent Heatless Dryer Cyclin...

Страница 85: ...Procedure Section 8 000 16 000 24 000 32 000 40 000 Install 8 000 Hour Maintenance Kit P011813 11 6 Install 16 000 Hour Maintenance Kit P011814 11 6 Read Record Flow Rate FLOW 8 4 4 1 Measure Record...

Страница 86: ...cessories 11 1 Top Section Parts Description Part Number Quantity Recommend Spare Solid State Relay P010562 2 1 Terminal Block P010200 1 Front Panel Display Assembly P017578 1 LCD Display Only P012105...

Страница 87: ...mmend Spare Boost Transformer P012772 1 Heatless Dryer Assembly See section11 3 for detail Pressure Gauge 0 30 PSI P011339 1 Static Pressure Regulator P013203 1 1 Adapter Plate 2 Flow Block 1 Air Tank...

Страница 88: ...Parts Description Part Number Quantity Recommend Spare Heatless Dryer P010196 1 Pressure Gauge P010695 1 Solenoid Valve In Kit P011813 See section 11 6 for detail Unloader Valve P010453 1 Cycle Timer...

Страница 89: ...T In Kit P011766 See section 11 6 for detail Air Compressor Assembly Left Side Right Side P013263 P013264 1 1 Compressor ONLY P010444 1 1 Spacer 8 Precooler P05663 2 Discharge Hose P05069 2 Capacitor...

Страница 90: ...P017585 Rev A 11 5 Frame Section Parts Description Part Number Quantity Recommend Spare Top Cover 1 Front Panel 1 Air Intake Filter In Kit P011766 See section 11 6 for detail Purge Muffler In Kit P011...

Страница 91: ...nth Maintenance Kit Includes air intake filter compressor mufflers and purge mufflers P011766 2 8 000 Hour Maintenance Kit Includes heatless dryer maintenance kits and compressor maintenance kits P011...

Страница 92: ...t the ALTEC AIR Inside Sales Service department to order 800 521 5351 option 2 Fax 303 657 2205 sales AltecAIR com parts AltecAIR com IMPORTANT Instruction for the replacement of individual listed com...

Страница 93: ...place desiccant o rings check valves springs and complete solenoid assembly o Test proper component operation Desiccant Tower Repack o Clean out tower and replace desiccant filter and o ring Circuit B...

Страница 94: ...begins with the most likely cause s of the issue Otherwise they start from the simplest possibilities and progress to more complicated ones This troubleshooting guide is designed to be easy to follow...

Страница 95: ...ng the air dryer may be necessary before performing certain procedures NEVER remove pressure sensing tubes from the Control Board without depressurizing the air dryer first or damage to the Control Bo...

Страница 96: ...running from the Control Board to the Display Screen is properly connected at both ends see section 11 1 for Control Board and Display Screen locations Plug in ribbon cable to Control Board and Displa...

Страница 97: ...UTP reading is higher than the High Outlet Pressure Alarm set point above Replace Control Board section 11 1 if Outlet Pressure OUTP reading is over verified High Outlet Pressure Alarm set point for m...

Страница 98: ...ntrol Board section 11 1 if Outlet Pressure OUTP reading is under verified Low Outlet Pressure Alarm set point for more than 1 minute and fails to create an alarm 13 9 High Flow Rate Alarm Possible Ca...

Страница 99: ...h Humidity Alarm set point section 9 13 Over 10 not recommended Heatless Dryer not cycling between towers Verify consistent Heatless Dryer cycling section 9 6 Troubleshoot Inconsistent Heatless Dryer...

Страница 100: ...ce Control Board if no alarm is created and system does not shut down section 11 1 13 13 Compressor Doesn t Operate Possible Cause Check Corrective Action Defective compressor Measure voltage to compr...

Страница 101: ...s between compressor and Air Tank for air leaks section 9 16 Connect tighten or replace leaking component 13 15 Compressor Excessive AMP Draw Possible Cause Check Corrective Action Restriction in air...

Страница 102: ...Alarm Possible Cause Check Corrective Action Low System Pressure Verify System Pressure section 8 12 Adjust System Pressure to 50 PSI section 8 12 High Flow condition Verify Flow Rate FLOW reading is...

Страница 103: ...set point by 1 minute or more and fails to create an alarm 13 19 Compressor Rapid ON OFF Cycling Possible Cause Check Corrective Action Defective Solid State Relay Measure voltages at Solid State Rel...

Страница 104: ...sted all of the potential problems and solutions covered in the Troubleshooting Your Dryer section and you still require further assistance to correct a problem contact ALTEC AIR Technical Support 800...

Страница 105: ...Altec AIR P8400W Air dryer User s Guide Page 105 of 110 P017585 Rev A 14 Appendix 14 1 Wiring Diagram...

Страница 106: ...Defined by Customer Device Model Alphanumeric Factory Preset Device Firmware Version Numeric Factory Preset Current Date Time Numeric mm dd yy hh mm IP Address Numeric xxx xxx xxx xxx Subnet Mask Num...

Страница 107: ...Flow Alarm OK Alarm High Outlet Pressure Alarm OK Alarm Low Outlet Pressure Alarm OK Alarm High Humidity System 1 Alarm OK Alarm High Humidity System 2 Alarm OK Alarm High Cabinet Temperature Alarm O...

Страница 108: ...the user and the user in turn should file any damage claim against the carrier All warranty items are F O B Broomfield Colorado Freight charges are the responsibility of the user This warranty shall n...

Страница 109: ...do 80020 800 521 5351 303 427 3700 Fax 303 657 2233 info AltecAir com www AltecAir com 16 2 Sales 800 521 5351 option 2 Fax 303 657 2205 sales AltecAir com parts AltecAir com 16 3 Service 800 521 5351...

Страница 110: ..._____________________________ ________________________________________________________________________ ________________________________________________________________________ ________________________...

Отзывы: