Installation and setup
Page C-12
Operator’s Manual PiCCO
2
Version 3.1
3.4
Enter patient data
Figure 17: Start screen
Monitoring of the previous patient can be continued. In this case all previous measurement results will be
stored by the monitor. PCCO calibration can be restored up to 15 minutes after the last shutdown of the
device. ScvO
2
calibration can be restored up to 24 hours after the last calibration.
CAUTION:
When restoring the calibration in the PCCO and ScvO
2
calibration menu an old
calibration factor may be used. Ensure that the values are plausible after having used this option.