PULSAR MEASUREMENT
117
P*102 Material
This parameter should be set to the type of material being monitored.
OPTION
DESCRIPTION
1 = Liquid (Default)
Used for liquids and flat solid materials.
2 = Solid
Used for sold material that is heaped or at an angle.
3 = Closed Tank
Use for applications within a closed tank or where a
secondary echo response may become focused to
create a larger echo than the first.
Dimensions
P*105 Empty Level
This parameter is to be set to the
maximum distance
from the
face
of the
transducer to the
empty point
, in
P*104
Measurement
Units
. Note this
value affects span as well, (see the following important information notices),
so should be set before span.
P*106 Span
This parameter should be set to the maximum distance from the
Empty
Level
(
P*105
) to the maximum material level. It is automatically set to be
equal to the
Empty
Level
(
P*105
) less the
Near
Blanking
distance (
P*107
)
when you set the empty level.
Important Notice
When changing the Empty Distance (P*105) you can also recalculate the
values for the Span so that it equals the empty distance (P*105) minus
Near Blanking (P*107) and the Relay Setpoints, so that they remain at the
same percentage values of the empty distance as they were before you
changed the empty distance (P*105). You will be asked the question
“Recalculate Span?” if you choose yes (enter 1), then the span will be
recalculated. Any other answer will leave the span at its original value.
You will then be asked if you want to “Recalculate Setpoints?”, if you
choose yes (enter 1), then all Relay Setpoints will be recalculated as a
percentage of the new empty distance. Any other answer will leave the
setpoints at their original values.
Important Notice
When using the dB Mach 3 the empty distance is measured from the end of the
horn to the empty point in P*104 Measurement Units.