Page
36
Trace Tab
Upon opening the trace tab, the information at the bottom of the screen will scroll
through
Measurement in progress, Updating mA, Updating
Trace. Once the trace
has updated the diagnostic traces from the sensor will be displayed. You are also
able to see the current velocity measurement obtained by the sensor.
The vertical axis is related to the signal strength, while the horizontal axis is related
to the detected flow.
When a signal is received, the ‘
green square
indicator
’ will move down from 10
on the vertical axis. The further the ‘
green square indicator
’ moves from 10, the
larger the strength of the received signal. During flow condition, the ‘
green
square indicator
’ will normally be within the range of 30 to 70. Under no-flow
conditions the ‘
green square indicator
’ will remain at 10.
When flow is detected, the ‘
red circle indicator
’ will move horizontally to
indicate flow detection. The ‘
red circle indicator
’ should stay at 0 under no-flow
conditions, and move along the horizontal axis when flow is detected.
Under flow condition, the line should still approach and flatten out at just below
the signal index 10 line. If the trace remains flat or the indicators struggle to
remain stable during flow, then increase the ‘Gain’ (P102) setting.
Содержание MicroFlow-i
Страница 1: ...MICROFLOW INSTRUCTION MANUAL...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 12: ...Page 6 EU Declaration of Conformity...
Страница 27: ...Page 21 Diagram 4 Diagram 3 Distance Distance Distance Distance Distance...
Страница 32: ...Page 26 This page is left blank intentionally...
Страница 34: ...Page 28 This page is left blank intentionally...