FLOWPULSE INSTRUCTION MANUAL
60
CHAPTER 7 TROUBLESHOOTING
LED Indications
GREEN
RED
STATUS
Flashing
On
Normal operation, with alarm condition
Flashing
Off
Normal operation, no alarm condition
Flashing
Off with Alarm
condition
A recoverable fault has occurred, check
Fault Codes (query registers 270
–
279) If
repetitive, Pulsar servicing maybe
required
Off
Flashing
In Bootloader mode/Waiting for
connection from Bootloader PC software.
Off
Off
Device not powered, or if powered, a
critical fault has developed.
Fault Registers and Log File
The registers 270-279 can be queried (via FlowPulse PC or Modbus) to check
for fault codes:
1.
Retrieved parameter exceeded valid range.
2.
Parameter value entered by user exceeded valid range.
3.
Number of self-reset events (register 266) exceeded the maximum
allowed.
4.
Fail to retrieve stored parameters.
While using FlowPulse PC, all parameter changes are time-stamped and
logged to a file within the Debug folder.