Runtime Registers (Input Registers)
The first table below presents the unit symbols and their corresponding
descriptions for the
Runtime Registers
. The second table shows the
addresses, descriptions, and units of the
Runtime Registers
. These registers
are ‘Read Only’ and can be accessed using the Function Code 4.
Example:
You wish to know the current value for Distance.
Search for register address 4 using the Function code 4 and the information
will then be displayed.
UNIT
DEFINITION
U_NO
None
U_MU
Measurement Unit
U_PC
Percentage
U_VU
Volumetric Unit
V_TP
Degree C
U_MN
Minutes
32BIT
REGISTER
ADDRESS
2X 16BIT
REGISTER
ADDRESS
DESCRIPTION
UNIT
000
2000-2001
Level
U_MU
001
2002-2003
Echo Status
U_NO
002
2004-2005
Echo strength
U_PC
003
2006-2007
Temperature
U_TP
004
2008-2009
Distance
U_MU
005
2010-2011
Volume
U_MN
006
2012-2013
Space
U_MU
007
2014-2015
Level/Distance/Space or Volume
(dependent on what P100 is set to)
U_MU
or U_VU
016
2032-2033
Minimum Temperature
U_TP
017
2034-2035
Maximum Temperature
U_TP
200
*2200
Transducer Type (P101)
U_NO
*This 16bit integer register is formatted as big endian (big endian bytes in
the register).