IMP LITE INSTRUCTION MANUAL
36
CHAPTER 6 TROUBLESHOOTING
This section describes many common symptoms, with suggestions as to
what to do.
SYMPTOM
WHAT TO DO
Display blank, transducer
not firing.
Check power supply
Display shows LOE
No valid echo being received, and unit has
gone into fault condition. Check material
level is not out of range, sensor is
perpendicular to material surface.
Incorrect reading being
displayed for current level
Measure actual distance from transducer face
to surface of material. Enter Program Mode
and access
SyS1
menu, go to
P21
and press
ENTER
type in the
measured distance
, press
ENTER
,
wait until “
Stor
” is displayed and
return to Run Mode, display should now
update to correct reading.
Material level is consistently
incorrect by the same
amount.
Check empty level (P105) correctly entered.