FLOWCERT LITE (UL) INSTRUCTION MANUAL
52
Example 1 V-Notch Weir
In this example it is required to calculate the flow through a Simple
Exponential Device, which on this occasion is a V-Notch Weir. The K factor
for the weir is unknown so ratiometric calculation will be used, there is no
requirement for alarms and the flow rate is to be displayed in litres/second.
The totaliser is to record the flow in cubic metres but is not to be displayed
during RUN.
The distance from the face of the lower transducer (DUET) to
zero
flow
(
P105
) is 1 metre and
max head
(
P704
) is 0.4 metres,
maximum
flow
(
P705
) is known to be 2.2 Million US gallons/day.
Transducer mounted
Min. blanking distance (P107)
above max. head
h
Minimum
3 x hmax
TYPICAL SIMPLE EXPONENTIAL DEVICE
DRWG. 5