Product User Guide | 2
MicroRHTemp
PRODUCT USER GUIDE
Product Overview
The MicroRHTemp is a battery powered, stand alone
humidity and temperature recorder that can fit in the
tightest places. It is even small enough to fit into most pill
bottles. It features an LED alarm indicator that alerts when
user-chosen temperature or humidity limits are exceeded.
This all-in-one compact, portable, easy to use device
will measure and record up to 16,383 measurements per
channel. The storage medium is non-volatile solid state
memory, providing maximum data security even if the
battery becomes discharged. Start and stop the device
directly from your computer or with programmable start
time and data retrieval is quick and easy. Simply plug it
into an empty USB port with an interface cable (IFC202)
and our user-friendly software does the rest.
Installation Guide
Installing the Software
The Software can be downloaded from the MadgeTech
website at
. Follow the instructions
provided in the Installation Wizard.
Installing the Interface Cable
IFC202 (sold separately)
— Connect the device into a USB
port with the interface cable and install the drivers.
Ordering Information
• 900362-00 — MicroRHTemp
• 900309-00 — IFC202
• 901570-00 — SR1154W Replacement battery (2 required)
Device Operation
Connecting and Starting the Data Logger
1. Once the software is installed and running, plug the
interface cable into the data logger.
2. Connect the USB end of the interface cable into an
open USB port on the computer.
3. The data logger will automatically appear under
Connected Devices
within the software.
4. For most applications, select
Custom Start
from
the menu bar and choose the desired start method,
reading rate and other parameters appropriate for the
data logging application and click
Start
.
(
Quick Start
applies the most recent custom start options,
Batch
Start
is used for managing multiple loggers at once,
Real Time Start
stores the dataset as it records while
connected to the logger.)
5. The status of the device will change to
Running
or
Waiting to Start
, depending upon your start method.
6. Disconnect the data logger from the interface cable
and place it in the environment to measure.
Note:
The device will stop recording data when the end of memory is
reached or the device is stopped, unless user selectable memory wrap
is enabled. At this point the device cannot be restarted until it has been
re-armed by the computer.
Downloading Data from a Data Logger
1. Connect the data logger to the computer with the
interface cable.
2. Highlight the data logger in the
Connected Device
s
list. Click
Stop
on the menu bar.
3. Once the data logger is stopped, with the logger
highlighted, click
Download
.
4. Downloading will offload and save all the recorded
data to the PC.